beautypg.com

Quick-start operating instructions – Dynasonics TFXM Series Ultrasonic Multi-Channel Flow Meter User Manual

Page 4

background image

Rev. 8/02

-1. 3-

TFXM

This manual contains detailed operating instructions for all
aspects of the TFXM instrument. The following condensed
instructions are provided to assist the operator in getting the
instrument started up and running as quickly as possible. This
pertains to basic operation only. If specific instrument features
are to be used or if the installer is unfamiliar with this type of
instrument, refer to the appropriate section in the manual for
complete details.

1. TRANSDUCER LOCATION

A. In general, select a mounting location on the piping system

with a minimum of 10 pipe diameters (10 X the pipe inside
diameter) of straight pipe upstream and 5 straight
diameters downsteam. The installation location should
also be positioned so that the pipe remains full when the
liquid is flowing through it. On horizontal pipes the trans-
ducers should be located on the sides of the pipe. See
Figure 1.2
. See Table 2.1 for additional configurations.

B. Select a mounting method, Figure 1.1, for the transducers

from Table 2.2, based on pipe size and liquid
characteristics. In General, select W-Mount for plastic and
steel pipes flowing clean, non-aerated liquids in the 1-6
inch [25-150 mm] internal diameter range. Select V-Mount
for pipes of all materials and most liquids in pipe sizes from
3-10 inches [75-400 mm]. Select Z-Mount for pipes larger
than 10 inches [400 mm].

C. For each measuring channel integrated into the TFXM, en-

ter the parameters listed in Table 1.1 via the TFXM keypad
or UltraLink software utility.

D. Record the value calculated and displayed as Transducer

Spacing/XDCR SPC.

Transducer
Location

QUICK-START OPERATING INSTRUCTIONS

W-Mount

V-Mount

Z-Mount

Figure 1.1