Multichannel operation – Dynasonics TFXM Series Ultrasonic Multi-Channel Flow Meter User Manual
Page 84

Rev. 8/02
-5. 2-
TFXM
highlighting to the next line of the display. Pressing the (soft)
SELECT key on the bottom line of the display will cause the
display configuration to end and the setup will be saved.
Multiple Pipe Flow Network
The TFXM is designed to be the Primary (Master) flow controller
in a network of Secondary (Slave) flow meters. The TFXM can be
used to display flow information and configure up to 119 satellite
flow meters connected on its RS485, two-wire network.
To configure a multiple pipe setup, follow the entries shown in
Figure 5.1. The Primary board in the TFXM is established as the
Master and all of the Secondary boards are considered slaves.
The coefficient values located on the bottom of the window do
not have a bearing on the display, because manipulated display
MULTICHANNEL OPERATION
Multiple Pipe
Configuration