Part 3 - keypad configuration – Dynasonics TFXM Series Ultrasonic Multi-Channel Flow Meter User Manual

Page 64

background image


Rev. 8/02

-3. 26-

TFXM

Signal Strength Cutoff

Signal Strength Cutoff SIG C-OF is used to drive the flowmeter
and its outputs to a zero flow state should conditions occur that
cause low signal strength. A signal strength indication of be-
tween 0.5 and 0.8 is considered to be inadequate for measur-
ing flow reliably, so typical settings for SIG C-OF are in the
range of 1.0 to 2.0.

Substitute Flow

Substitute Flow or SUB FLOW is a value that the analog out-
puts will be driven at when an error condition in the flowmeter
occurs. Typical settings are either –5% or 105% - a value out-
side of the normal operating range that can be used to indicate
a fault condition to the target device.

PART 3 - KEYPAD CONFIGURATION

Signal Strength
Cutoff

MIN RATE
SETTING

MAX RATE

SETTING

SUB FLOW

SETTING

DISPLAY

READING READING

DURING ERRORS

0.0

1,000.0

0.0

0.000

-500.0

500.00

50.0

0.000

-100.0

200.0

33.3

0.000

0.0

1,000.0

-5.0*

-50.00

Substitute Flow
Entry

Signal Strength indication of 0.5 to 0.8 is considered to be no
signal at all. Verify that the pipe if full of liquid, the pipe size and
liquid parameters are entered correctly and that the transducers
have been mounted accurately.