beautypg.com

Atec Agilent-PNA-X Series User Manual

Page 8

background image

8

Table 1. System Dynamic Range, Options 200 or 400, and Receiver Dynamic Range, All Options

Description

Specification, Options 200, 400

Typical, All Options

System

Dynamic

Range (dB)

(A)-(B)

Max Leveled

Output

Power
(dBm)

(A)

Test Port

Noise

Floor

(dBm)

(B)

Receiver
Dynamic

Range (dB)

(C)-(D)

Test Port

Compression

at 0.1 dB

(dBm)

(C)

Test Port

Noise

Floor

(dBm)

(D)

Ports 1,3

1

Ports 1,3

1

All Ports

All Ports

All Ports

All Ports

10 MHz to 50 MHz

93

13

-80

--

--

-87

50 MHz to 100 MHz

103

13

-90

--

--

-95

100 MHz to 500 MHz

117

13

-104

125

15

-110

500 MHz to 3.2 GHz

124

10

-114

130

13

-117

3.2 GHz to 8.5 GHz

127

13

-114

130

13

-117

8.5 GHz to 13.5 GHz

127

13

-114

130

13

-117

13.5 GHz to 16 GHz

127

13

-114

130

13

-117

16 GHz to 20 GHz

127

13

-114

129

12

-117

20 GHz to 24 GHz

122

12

-110

125.5

10.5

-115

24 GHz to 26.5 GHz

112

5

-107

123

10

-113

1

Either port can be used as the source port. Any other port can be used as the receiver port.

Table 2a. System Dynamic Range at Test Port (dB), Options 200 or 400

Description

Specification

Typical

Ports 1, 3

1

Ports 2, 4

1

Ports 1, 3

1

Ports 2, 4

1

10 MHz to 50 MHz

93

93

106

104

50 MHz to 100 MHz

103

103

116

115

100 MHz to 500 MHz

117

117

131

130

500 MHz to 3.2 GHz

124

127

130

135

3.2 GHz to 8.5 GHz

127

127

137

136

8.5 GHz to 10 GHz

127

127

137

136

10 GHz to 13.5 GHz

127

127

134

133

13.5 GHz to 16 GHz

127

127

134

133

16 GHz to 20 GHz

127

124

133

129

20 GHz to 24 GHz

122

117

130

126

24 GHz to 26.5 GHz

112

109

124

120

1