Atec Agilent-PNA-X Series User Manual
Page 8
8
Table 1. System Dynamic Range, Options 200 or 400, and Receiver Dynamic Range, All Options
Description
Specification, Options 200, 400
Typical, All Options
System
Dynamic
Range (dB)
(A)-(B)
Max Leveled
Output
Power
(dBm)
(A)
Test Port
Noise
Floor
(dBm)
(B)
Receiver
Dynamic
Range (dB)
(C)-(D)
Test Port
Compression
at 0.1 dB
(dBm)
(C)
Test Port
Noise
Floor
(dBm)
(D)
Ports 1,3
1
Ports 1,3
1
All Ports
All Ports
All Ports
All Ports
10 MHz to 50 MHz
93
13
-80
--
--
-87
50 MHz to 100 MHz
103
13
-90
--
--
-95
100 MHz to 500 MHz
117
13
-104
125
15
-110
500 MHz to 3.2 GHz
124
10
-114
130
13
-117
3.2 GHz to 8.5 GHz
127
13
-114
130
13
-117
8.5 GHz to 13.5 GHz
127
13
-114
130
13
-117
13.5 GHz to 16 GHz
127
13
-114
130
13
-117
16 GHz to 20 GHz
127
13
-114
129
12
-117
20 GHz to 24 GHz
122
12
-110
125.5
10.5
-115
24 GHz to 26.5 GHz
112
5
-107
123
10
-113
1
Either port can be used as the source port. Any other port can be used as the receiver port.
Table 2a. System Dynamic Range at Test Port (dB), Options 200 or 400
Description
Specification
Typical
Ports 1, 3
1
Ports 2, 4
1
Ports 1, 3
1
Ports 2, 4
1
10 MHz to 50 MHz
93
93
106
104
50 MHz to 100 MHz
103
103
116
115
100 MHz to 500 MHz
117
117
131
130
500 MHz to 3.2 GHz
124
127
130
135
3.2 GHz to 8.5 GHz
127
127
137
136
8.5 GHz to 10 GHz
127
127
137
136
10 GHz to 13.5 GHz
127
127
134
133
13.5 GHz to 16 GHz
127
127
134
133
16 GHz to 20 GHz
127
124
133
129
20 GHz to 24 GHz
122
117
130
126
24 GHz to 26.5 GHz
112
109
124
120
1