Table 34. (Continued) Rear Panel Information, All Options
Description
Typical Performance
External IF Inputs
Function
Allows use of external IF signals from remote mixers, bypassing the PNA's first converters
Connectors
SMA (female); A, B, C, D, R (4-port); A, B, R1, R2 (2-port)
Frequency
Path
Normal IF path:
Narrowband IF path:
DSP Version
4
5
4 or 5
IF Bandwidth
All
All
<= 600 kHz
1 MHz
1.5 MHz
2 MHz
3 MHz
5 MHz
7 MHz
10 MHz
15 MHz
All
RF Frequency
< 53 MHz
>= 53 MHz
< 53 MHz
>= 53 MHz
All
All
All
All
All
All
All
All
All
IF Frequency
2.535211 MHz
7.605634 MHz
2.479339 MHz
7.438017 MHz
7.692 MHz
7.368 MHz
8.450 MHz
8.163 MHz
6.897 MHz
10.53 MHz
15.38 MHz
22.22 MHz
10.70 MHz
Input Impedance
50
RF Damage Level
+23 dBm
DC Damage Level
5.5 VDC
0.1 dB Compression Point
Normal IF path
Narrowband IF path
-9.0 dBm at 7.438 MHz
-17 dBm at 10.70 MHz
Pulse Inputs (IF Gates)
Function
Internal receiver gates used for point-in-pulse and pulse-profile measurements
Connectors
15-pin mini D-sub
Input Impedance
1 K Ohm
Source Modulators
20 ns minimum pulse width
Receiver Gates
20 ns minimum pulse width
DC Damage Level
5.5 VDC
Drive Voltage
0 V (off), +3.3 V (on), nominal