Table 22. reference level - specification, Table 21a. trace noise, Magnitude (db rms), all options, all ports – Atec Agilent-PNA-X Series User Manual
Page 33: Table 21b. trace noise, Phase (deg rms), all options, all ports
33
16 GHz to 24 GHz
<0.23
24 GHz to 26.5 GHz
<0.29
1
Test port receiver compression at specified input levels below 500 MHz is negligible due to coupler roll.
Table 21a. Trace Noise
1
Magnitude (dB rms), All Options, All Ports
Description
Specification
Typical
1 kHz IFBW
1 kHz IFBW
100 kHz IFBW
600 kHz IFBW
10 MHz to 100 MHz
0.007
0.0039
0.040
0.140
100 MHz to 8.5 GHz
0.002
0.0005
0.005
0.011
8.5 GHz to 13.5 GHz
0.002
0.0005
0.005
0.011
13.5 GHz to 16 GHz
0.002
0.0005
0.005
0.011
16 GHz to 22.5 GHz
0.002
0.0006
0.005
0.012
22.5 GHz to 24 GHz
0.003
0.0014
0.008
0.020
24 GHz to 26.5 GHz
0.005
0.0020
0.008
0.020
1
Ratioed measurement, nominal power at test port.
Table 21b. Trace Noise
1
Phase (deg rms), All Options, All Ports
Description
Specification
Typical
1 kHz IFBW
1 kHz IFBW
100 kHz IFBW
600 kHz IFBW
10 MHz to 100 MHz
0.051
0.0261
0.266
1.053
100 MHz to 8.5 GHz
0.015
0.0041
0.030
0.075
8.5 GHz to 13.5 GHz
0.015
0.0041
0.030
0.075
13.5 GHz to 16 GHz
0.042
0.0124
0.030
0.075
16 GHz to 22.5 GHz
0.042
0.0135
0.033
0.082
22.5 GHz to 26.5 GHz
0.054
0.0225
0.057
0.139
1
Ratioed measurement, nominal power at test port.
Table 22. Reference Level - Specification
Description
Magnitude (dB)
Phase (°)
Range
+/-500
+/-500
Resolution
0.001
0.01