Dynamic Accuracy

Table 30. Dynamic Accuracy - Specification

Standard receiver accuracy of the test port input power reading relative to the reference input power level. It is verified
with the following measurements:

 Compression over frequency

 IF linearity at a single frequency of 1.998765 GHz using a reference level of -20 dBm for an input power range

of 0 to -60 dBm. For value below -60 dBm, refer to “VNA Receiver Dynamic Accuracy Specifications and
Uncertainties”.

Applies to N5249A (all serial numbers) and N5241A, N5242A with following serial numbers:

N5241A and N5241AS: MY5241/SG5241/US5241 and above

N5242A and N5242AS: MY5242/SG5242/US5242 and above

Please download our free Uncertainty Calculator from

http://www.agilent.com/find/na_calculator

to generate the

curves for your PNA.

Dynamic Accuracy, 0.010 GHz

Dynamic Accuracy, 0.050 GHz