Atec Agilent-PNA-X Series User Manual

Page 14

background image

14

Table 5. N4433A 4-Port Electronic Calibration Module

Note: Uncertainty curves for the N4433A are created using a 2-port calibration. Multiport uncertainties are not
supported at this time.

Description

Specification (dB)

10 MHz to

50 MHz

50 MHz to

500 MHz

500 MHz to

2 GHz

2 GHz to

8.5 GHz

8.5 GHz to

13.5 GHz

13.5 GHz to

20 GHz

Directivity

50

50

50

45

45

40

Source
Match

42

42

42

37

37

31

Load Match

40

41

41

35

35

29

Reflection Tracking

1

Mag

±0.060

±0.060

±0.060

±0.100

±0.100

±0.180

Phase (°)

±0.396

±0.396

±0.396

±0.660

±0.660

±1.188

Transmission Tracking without Option 029

Mag

±0.068

±0.064

±0.064

±0.115

±0.115

±0.210

Phase (°)

±0.447

±0.421

±0.421

±0.761

±0.761

±1.387

S

21

Transmission Tracking with Option 029

1

Mag

±0.070

±0.066

±0.070

±0.126

±0.126

±0.225

Phase (°)

±0.460

±0.436

±0.460

±0.829

±0.829

±1.485

S

12

Transmission Tracking with Option 029

1

Mag

±0.082

±0.071

±0.072

±0.123

±0.123

±0.227

Phase (°)

±0.541

±0.469

±0.472

±0.812

±0.812

±1.496

The port 1 noise tuner switch set to the bypass position, and port 2 noise receiver switch set to the normal position.