Atec Agilent-PNA-X Series User Manual
Page 14
14
Table 5. N4433A 4-Port Electronic Calibration Module
Note: Uncertainty curves for the N4433A are created using a 2-port calibration. Multiport uncertainties are not
supported at this time.
Description
Specification (dB)
10 MHz to
50 MHz
50 MHz to
500 MHz
500 MHz to
2 GHz
2 GHz to
8.5 GHz
8.5 GHz to
13.5 GHz
13.5 GHz to
20 GHz
Directivity
50
50
50
45
45
40
Source
Match
42
42
42
37
37
31
Load Match
40
41
41
35
35
29
Reflection Tracking
1
Mag
±0.060
±0.060
±0.060
±0.100
±0.100
±0.180
Phase (°)
±0.396
±0.396
±0.396
±0.660
±0.660
±1.188
Transmission Tracking without Option 029
Mag
±0.068
±0.064
±0.064
±0.115
±0.115
±0.210
Phase (°)
±0.447
±0.421
±0.421
±0.761
±0.761
±1.387
S
21
Transmission Tracking with Option 029
1
Mag
±0.070
±0.066
±0.070
±0.126
±0.126
±0.225
Phase (°)
±0.460
±0.436
±0.460
±0.829
±0.829
±1.485
S
12
Transmission Tracking with Option 029
1
Mag
±0.082
±0.071
±0.072
±0.123
±0.123
±0.227
Phase (°)
±0.541
±0.469
±0.472
±0.812
±0.812
±1.496
The port 1 noise tuner switch set to the bypass position, and port 2 noise receiver switch set to the normal position.