beautypg.com

Combine, Mode - typical, All options – Atec Agilent-PNA-X Series User Manual

Page 25

background image

25

Table 9h. Maximum Leveled Power (dBm), Options 224 or 423 with 029

1

, Combine Mode - Typical

Description

Source 1 Port 1

Source 2 Port 1

Filtered Mode

2

Hi Power Mode

2

Filtered Mode

2

Hi Power Mode

2

10 MHz to 50 MHz

6

16

-8

2

50 MHz to 500 MHz

9

17

-5

4

500 MHz to 3.2 GHz

9

10

-5

-4

3.2 GHz to 8.5 GHz

15

15

2

2

8.5 GHz to 10 GHz

15

15

2

2

10 GHz to 13.5 GHz

11

11

-2

-2

13.5 GHz to 16 GHz

11

11

-2

-2

16 GHz to 20 GHz

8

8

-4

-4

20 GHz to 24 GHz

5

5

-7

-7

24 GHz to 26.5 GHz

0

0

-11

-11

1

Option 029 affects port 1 and port 2 maximum leveled power.

2

In Filtered Mode, the signal path goes through filters to minimize harmonics below 3.2 GHz. In Hi Power Mode, the

signal bypasses the filters to maximize output power.

Table 10. Power Level Accuracy (dB) at Nominal Power

1

, All Options

Description

Specification

Typical

Ports 1, 2, 3, 4

2

Source 2 Out 1
Source 2 Out 2

Ports 1, 2, 3, 4

2

Source 2 Out 1
Source 2 Out 2

10 MHz to 50 MHz

+/-1.0

+/-2.0

+/-0.40

+/-0.55

50 MHz to 500 MHz

+/-1.0

+/-2.0

+/-0.20

+/-0.25

500 MHz to 3.2 GHz

+/-1.0

+/-2.0

+/-0.25

+/-0.25

3.2 GHz to 8.5 GHz

+/-1.0

+/-2.0

+/-0.40

+/-0.25

8.5 GHz to 10 GHz

+/-1.0

+/-2.0

+/-0.40

+/-0.25

10 GHz to 13.5 GHz

+/-1.2

+/-2.0

+/-0.60

+/-0.25

13.5 GHz to 18 GHz

+/-2.0

+/-2.5

+/-0.60

+/-1.00

18 GHz to 26.5 GHz

+/-2.5

+/-2.5

+/-0.80

+/-0.90

1

Level accuracy at power other than nominal power, Power Level Accuracy (dB) at Nominal Power + Power Level

Linearity (dB)

2

Any port can be used as the source port. Source in filtered mode where applicable.