Table 4. 85052b calibration kit, The corrected system performance with cal kits – Atec Agilent-PNA-X Series User Manual
Page 12
12
N5241A, N5242A, and N5249A Corrected System Performance with 3.5mm Connectors, All Opts
Note: For any Sii reflection measurement:
Sjj = 0.
For any Sij transmission measurement:
Sji = Sij when Sij
1
Sji = 1/Sij when Sij
1
Skk = 0 for all k
Applies to the N5241A, N5242A, and N5249A Option 200 or 219 or 224 or 400 or 419 or 423 analyzers with serial
numbers listed below, 85131F flexible test port cable set, and a full 2-port calibration.
N5241A and N5241AS: MY5241/SG5241/US5241 and above
N5242A and N5242AS: MY5242/SG5242/US5242 and above
Also applies to the following condition: Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration
temperature. This document does not present specifications for the 85052C or 85052D Calibration Kit. The uncertainty
charts were generated for Option 423, although they are applicable for any option without Option 029. Please download
our free Uncertainty Calculator fro
calibration kit and PNA setup.
Table 4. 85052B Calibration Kit
Description
Specification (dB)
10 MHz to
50 MHz
50 MHz to
500 MHz
500 MHz to
2 GHz
2 GHz to
8.5 GHz
8.5 GHz to
13.5 GHz
13.5 GHz to
20 GHz
20 GHz to
26.5 GHz
Directivity
48
48
48
44
44
44
44
Source Match
40
40
40
31
31
31
31
Load Match
48
48
48
44
44
44
44
Reflection Tracking
1
Mag
±0.003
±0.003
±0.003
±0.006
±0.006
±0.006
±0.006
Phase (°)
±0.020
±0.020
±0.020
±0.040
±0.040
±0.040
±0.040
Transmission Tracking without Option 029
Mag
±0.034
±0.034
±0.017
±0.091
±0.091
±0.104
±0.119
Phase (°)
±0.225
±0.225
±0.110
±0.602
±0.602
±0.688
±0.788
S
21
Transmission Tracking with Option 029
1
Mag
±0.034
±0.034
±0.017
±0.091
±0.091
±0.104
±0.119
Phase (°)
±0.225
±0.225
±0.110
±0.602
±0.602
±0.688
±0.788
S
12
Transmission Tracking with Option 029
1
Mag
±0.017
±0.017
±0.017
±0.091
±0.091
±0.104
±0.119
Phase (°)
±0.110
±0.110
±0.110
±0.602
±0.602
±0.688
±0.788
The port 1 noise tuner switch set to the bypass position, and port 2 noise receiver switch set to the normal position.