Test port input, Table 18. noise floor, Dbm) at 10 hz ifbw, all options, all ports – Atec Agilent-PNA-X Series User Manual
Page 32
32
Test Port Input
Table 18. Noise Floor
1
(dBm) at 10 Hz IFBW, All Options, All Ports
Description
Specification
Typical
Test Port
Direct Receiver
Access Input
Test Port
Direct Receiver
Access Input
10 MHz to 50 MHz
2
-80
--
-87
-130
50 MHz to 100 MHz
2
-90
--
-95
-128
100 MHz to 500 MHz
2
-104
--
-110
-132
500 MHz to 2 GHz
-114
--
-117
-133
2 GHz to 8.5 GHz
-114
--
-117
-129
8.5 GHz to 13.5 GHz
-114
--
-117
-129
13.5 GHz to 20 GHz
-114
--
-117
-129
20 GHz to 24 GHz
-110
--
-115
-122
24 GHz to 26.5 GHz
-107
--
-113
-119
1
Total average (rms) noise power calculated as the mean value of a linear magnitude trace expressed in dBm.
2
May typically be degraded at particular frequencies below 500 MHz due to spurious receiver residuals.
Table 19. 0.1 dB Compression at Test Port (dBm), All Options, All Ports - Typical
Description
Typical
10 MHz to 500 MHz
1
--
500 MHz to 8.5 GHz
13
8.5 GHz to 13.5 GHz
13
13.5 GHz to 16 GHz
13
16 GHz to 20 GHz
12
20 GHz to 24 GHz
10.5
24 GHz to 26.5 GHz
10
1
Test port receiver compression at specified input levels below 500 MHz is negligible due to coupler roll.
Table 20. Test Port Compression (dB) at 8 dBm Test Port Power, All Options, All Ports, - Specification
Description
Specification
10 MHz to 500 MHz
1
--
500 MHz to 8.5 GHz
<0.17
8.5 GHz to 13.5 GHz
<0.17
13.5 GHz to 16 GHz
<0.17