beautypg.com

Test port input, Table 18. noise floor, Dbm) at 10 hz ifbw, all options, all ports – Atec Agilent-PNA-X Series User Manual

Page 32

background image

32

Test Port Input

Table 18. Noise Floor

1

(dBm) at 10 Hz IFBW, All Options, All Ports

Description

Specification

Typical

Test Port

Direct Receiver

Access Input

Test Port

Direct Receiver

Access Input

10 MHz to 50 MHz

2

-80

--

-87

-130

50 MHz to 100 MHz

2

-90

--

-95

-128

100 MHz to 500 MHz

2

-104

--

-110

-132

500 MHz to 2 GHz

-114

--

-117

-133

2 GHz to 8.5 GHz

-114

--

-117

-129

8.5 GHz to 13.5 GHz

-114

--

-117

-129

13.5 GHz to 20 GHz

-114

--

-117

-129

20 GHz to 24 GHz

-110

--

-115

-122

24 GHz to 26.5 GHz

-107

--

-113

-119

1

Total average (rms) noise power calculated as the mean value of a linear magnitude trace expressed in dBm.

2

May typically be degraded at particular frequencies below 500 MHz due to spurious receiver residuals.

Table 19. 0.1 dB Compression at Test Port (dBm), All Options, All Ports - Typical

Description

Typical

10 MHz to 500 MHz

1

--

500 MHz to 8.5 GHz

13

8.5 GHz to 13.5 GHz

13

13.5 GHz to 16 GHz

13

16 GHz to 20 GHz

12

20 GHz to 24 GHz

10.5

24 GHz to 26.5 GHz

10

1

Test port receiver compression at specified input levels below 500 MHz is negligible due to coupler roll.

Table 20. Test Port Compression (dB) at 8 dBm Test Port Power, All Options, All Ports, - Specification

Description

Specification

10 MHz to 500 MHz

1

--

500 MHz to 8.5 GHz

<0.17

8.5 GHz to 13.5 GHz

<0.17

13.5 GHz to 16 GHz

<0.17