Table 43. coupler inputs (db) - typical, Table 44. damage level - typical, Coupler inputs – Atec Agilent-PNA-X Series User Manual
Page 54
54
3.2 GHz to 8.5 GHz
14
14
14
8.5 GHz to 10 GHz
14
14
14
10 GHz to 13.5 GHz
14
14
12
13.5 GHZ to 16 GHz
14
14
12
16 GHZ to 20 GHz
12
12
9
20 GHZ to 24 GHz
9
9
7
24 GHZ to 26.5 GHz
2
2
4
1
In Filtered Mode, the signal path goes through filters to minimize harmonics below 3.2 GHz. In Hi Power Mode, the
signal bypasses the filters to maximize output power.
Table 43. Coupler Inputs (dB) - Typical
(PORT 1, 2, 3, 4 CPLR THRU) Insertion Loss of Coupler Thru
Description
Options 200, 400
Options 219, 419, 224, 423
10 MHz to 50 MHz
0
-0.5
50 MHz to 500 MHz
-0.25
-0.75
500 MHz to 3.2 GHz
-0.5
-1.0
3.2 GHz to 8.5 GHz
-0.75
-1.25
8.5 GHz to 10 GHz
-0.75
-1.25
10 GHz to 13.5 GHz
-1.0
-1.75
13.5 GHZ to 16 GHz
-1.0
-1.75
16 GHZ to 20 GHz
-1.5
-2.25
20 GHZ to 24 GHz
-1.5
-2.5
24 GHZ to 26.5 GHz
-1.75
-2.5
Table 44. Damage Level - Typical
Description
RF (dBm)
DC (V)
RCVR A, B, C, D IN
15
0
RCVR R1 IN
15
7
RCVR R2, R3, R4 IN
15
15
REF 1 SOURCE OUT
15
7
REF 2, 3, 4 SOURCE OUT
30
0
PORT 1, 2, 3, 4 SOURCE OUT
15
0
PORT 1, 2, 3, 4 CPLR THRU
30
0
PORT 1, 2, 3, 4 CPLR ARM
30
40