Table 36b. Cycle Time (ms) for Full-Span Measurement Completion - Typical

10 MHz to 26.5 GHz

Number of Points

IF Bandwidth

201

401

1601

16001

600 kHz

Uncorrected

59

69

118

350

2-Port cal

125

147

244

707

10 kHz

Uncorrected

94

156

480

2333

2-Port cal

196

320

968

4674

1 kHz

Uncorrected

277

504

1873

17950

2-Port cal

561

1015

3756

35900

Table 37. Cycle Time vs. IF Bandwidth - Typical

Applies to the Preset condition (201 points, correction off) except for the following changes:

 CF = 10 GHz

 Span = 100 MHz

 Display off (add 21 ms for display on)

Cycle time includes sweep and retrace time.

Description

Typical Performance

IF Bandwidth (Hz)

Cycle Time (ms)

Trace Noise

600,000

5.00

0.009

100,000

6.84

0.003

30,000

11.6

0.002

10,000

29.0

0.001

3,000

71.8

0.0007

1,000

222

0.0004

300

640

0.0003

100

1826

0.0002

30

5982

<0.0002

10

17830

<0.0002

3

60000

<0.0002