beautypg.com

Measurement throughput summary – Atec Agilent-PNA-X Series User Manual

Page 47

background image

47

Measurement Throughput Summary

Typical Cycle Time for Measurement Completion

Cycle Time vs. IF Bandwidth

Cycle Time vs. Number of Points

Data Transfer Time

Cycle time Includes sweep time, retrace time and band-crossing time. Analyzer display turned off with DISPLAY:ENABLE
OFF. Add 21 ms for display on. Data for one trace (S

11

) measurement.

Table 36a. Cycle Time (ms) for Measurement Completion, All Options - Typical

Sweep
Range

IF
Bandwidth

Number of Points

201

401

1601

16001

7 GHz to
8 GHz

600 kHz

Uncorrected

6

6.3

9.6

56

2-Port cal

20

21

28

134

10 kHz

Uncorrected

36

53

200

1945

2-Port cal

80

115

405

3900

1 kHz

Uncorrected

227

444

1740

17000

2-Port cal

460

900

3484

34000

9 GHz to
10 GHz

600 kHz

Uncorrected

6

6.3

9.6

56

2-Port cal

20

21

28

134

10 kHz

Uncorrected

36

53

200

1945

2-Port cal

80

115

405

3900

1 kHz

Uncorrected

227

444

1740

17000

2-Port cal

460

900

3484

34000

10 GHz to
13.5 GHz

600 kHz

Uncorrected

26

33

54

85

2-Port cal

62

77

121

190

10 kHz

Uncorrected

70

118

273

1958

2-Port cal

149

245

553

3922

1 kHz

Uncorrected

236

459

1780

17300

2-Port cal

400

926

3565

34600

13.5 GHz
to 20 GHz

600 kHz

Uncorrected

26

33

54

85

2-Port cal

62

77

121

190

10 kHz

Uncorrected

70

118

273

1958

2-Port cal

149

245

553

3922

1 kHz

Uncorrected

236

459

1780

17300

2-Port cal

400

926

3565

34600