47
Measurement Throughput Summary
Typical Cycle Time for Measurement Completion
Cycle Time vs. Number of Points
Cycle time Includes sweep time, retrace time and band-crossing time. Analyzer display turned off with DISPLAY:ENABLE
OFF. Add 21 ms for display on. Data for one trace (S
11
) measurement.
Table 36a. Cycle Time (ms) for Measurement Completion, All Options - Typical
Sweep
Range
IF
Bandwidth
Number of Points
201
401
1601
16001
7 GHz to
8 GHz
600 kHz
Uncorrected
6
6.3
9.6
56
2-Port cal
20
21
28
134
10 kHz
Uncorrected
36
53
200
1945
2-Port cal
80
115
405
3900
1 kHz
Uncorrected
227
444
1740
17000
2-Port cal
460
900
3484
34000
9 GHz to
10 GHz
600 kHz
Uncorrected
6
6.3
9.6
56
2-Port cal
20
21
28
134
10 kHz
Uncorrected
36
53
200
1945
2-Port cal
80
115
405
3900
1 kHz
Uncorrected
227
444
1740
17000
2-Port cal
460
900
3484
34000
10 GHz to
13.5 GHz
600 kHz
Uncorrected
26
33
54
85
2-Port cal
62
77
121
190
10 kHz
Uncorrected
70
118
273
1958
2-Port cal
149
245
553
3922
1 kHz
Uncorrected
236
459
1780
17300
2-Port cal
400
926
3565
34600
13.5 GHz
to 20 GHz
600 kHz
Uncorrected
26
33
54
85
2-Port cal
62
77
121
190
10 kHz
Uncorrected
70
118
273
1958
2-Port cal
149
245
553
3922
1 kHz
Uncorrected
236
459
1780
17300
2-Port cal
400
926
3565
34600