beautypg.com

Uncorrected system performance, Table 7a. directivity (db) – Atec Agilent-PNA-X Series User Manual

Page 18

background image

18

Uncorrected System Performance

Specifications apply to following conditions:

Over environmental temperature of 25 °C ±5 °C, with less than 1°C variation from the calibration temperature.

Cable loss not included in Transmission Tracking.

Crosstalk measurement conditions: normalized to a thru, measured with shorts on all ports, 10 Hz IF
bandwidth, averaging factor of 8, alternate mode, source power set to the specified maximum power.

With option 029, port 1 impedance tuner switch is in external position and port 2 noise receiver switch is in
noise receiver position unless specified. Refer to Options 200, 219, 224, 400, 419, 423 for performance of Option
029 Port 1 with impedance tuner switch in internal position, Port 2 noise receiver switch in normal position,
Ports 3 and 4.

Table 7a. Directivity (dB)

Description

Specification

Typical

Options

200, 219, 224, 400,

419, 423

Option 029

Options

200, 219, 224, 400,

419, 423

Option 029

All Ports

Ports 1, 2

All Ports

Ports 1, 2

10 MHz to 50 MHz

16

16

23

22

50 MHz to 500 MHz

24

24

28

28

500 MHz to 3.2 GHz

24

24

32

32

3.2 GHz to 8.5 GHz

23

23

25

25

8.5 GHz to 10 GHz

23

23

25

25

10 GHz to 13.5 GHz

16

16

22

20

13.5 GHz to 16 GHz

16

16

22

20

16 GHz to 20 GHz

16

15

22

20

20 GHz to 24 GHz

16

15

22

20

24 GHz to 26.5 GHz

16

15

22

20