Test port output, Table 8. frequency information, all options – Atec Agilent-PNA-X Series User Manual
Page 21
21
Test Port Output
See Block diagrams for all models and options beginning on page 59.
With option 029, port 1 noise tuner switch is in internal position and port 2 noise receiver switch is in normal position
unless specified.
Table 8. Frequency Information, All Options
Description
Specification (dB)
Typical (dB)
N5249A Frequency Range
10 MHz to 8.5 GHz
--
N5241A Frequency Range
10 MHz to 13.5 GHz
--
N5242A Frequency Range
10 MHz to 26.5 GHz
--
Frequency Resolution
1 Hz
--
Frequency Accuracy
+/- 1 ppm
--
Frequency Stability
--
+/-0.05 ppm, -10° to 70° C
1
+/-0.1 ppm/yr maximum
2
1
Assumes no variation in time.
2
Assumes no variation in temperature.
Table 9a. Maximum Leveled Power (dBm), Options 200 or 400
Description
Specification
Typical
Ports 1, 3
1
Ports 2, 4
1
Ports 1, 3
1
Ports 2, 4
1
Filtered
Mode
2
Hi Power
Mode
2
Filtered
Mode
2
Hi Power
Mode
2
10 MHz to 50 MHz
8
13
13
10
19
17
50 MHz to 500 MHz
10
13
13
11
21
20
500 MHz to 3.2 GHz
10
10
13
12
13
18
3.2 GHz to 8.5 GHz
13
13
13
20
20
19
8.5 GHz to 10 GHz
13
13
13
20
20
19
10 GHz to 13.5 GHz
13
13
13
17
17
16
13.5 GHz to 16 GHz
13
13
13
17
17
16
16 GHz to 20 GHz
13
13
10
16
16
12
20 GHz to 24 GHz
12
12
7
15
15
11
24 GHz to 26.5 GHz
5
5
2
11
11
7
1
Either port can be used as the source port.
2
In Filtered Mode, the signal path goes through filters to minimize harmonics below 3.2 GHz. In Hi Power Mode, the
signal bypasses the filters to maximize output power.