Atec Agilent-E8364C User Manual
Page 9

9
E8362/3/4C
Corrected system performance with 3.5 mm connectors
Description Specification
(dB)
45 MHz to 2 GHz
2 to 20 GHz
20 to 26.5 GHz
Directivity 48
44
44
Source match
40
31
31
Load match
48
44
44
Reflection tracking
±0.003 (+0.02/°C)
±0.006 (+0.02/°C)
±0.006 (+0.03/°C)
Transmission tracking
±0.009 (+0.02/°C)
±0.088 (+0.02/°C)
±0.104 (+0.03/°C)
Transmission uncertainty (specifications)
Reflection uncertainty (specifications)
Magnitude
0.01
0.1
1
10
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Transmission coefficient (dB)
Uncertainty (dB)
E836xC full two port cal using 85052B
S11 = S22 = 0
Source power = -12 dBm
45 MHz to 2 GHz
2 to 20 GHz
20 to 26.5 GHz
Phase
0.1
1
10
100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Transmission coefficient (dB)
Uncertainty (degrees)
E836xC full two port cal using 85052B
S11 = S22 = 0
Source power = -12 dBm
45 MHz to 2 GHz
2 to 20 GHz
20 to 26.5 GHz
Magnitude
0
0.01
0.02
0.03
0.04
0.05
0
0.2
0.4
0.6
0.8
1
Reflection coefficient (linear)
Uncertainty (linear)
E836xC with 85052B
S21 = S12 = 0
Source power = -12 dBm
45 MHz to 2 GHz
2 to 20 GHz
20 to 26.5 GHz
Phase
0
2
4
6
8
10
0
0.2
0.4
0.6
0.8
1
Reflection coefficient (linear)
Uncertainty (deg)
E836xC with 85052B
S21 = S12 = 0
Source power = -12 dBm
45 MHz to 2 GHz
2 to 20 GHz
20 to 26.5 GHz
Standard configuration and standard power range (E836xC)
Applies to E836xC PNA Series analyzer, 85052B (3.5 mm) calibration kit, 85131F flexible test
port cable set, and a full two-port calibration. (Specifications apply over environmental temperature of
23 °C ±3 °C, with less than 1 °C deviation from calibration temperature.) Data and traces above 20 GHz
applicable to E8363C and E8364C network analyzers, when used with 85052B 26.5 GHz calibration kit.
Data and traces above 20 GHz not applicable to E8362C.