Microwave PNA Series

Measurement throughput summary

Cycle time vs. IF bandwidth 1

Cycle time vs. number of points 1

Instrument state: preset condition, 201 points,

Instrument state: preset condition, 35 kHz IF

CF = 28 GHz, Span = 100 MHz, correction off. Add

bandwidth, CF = 28 GHz, Span = 100 MHz,

21 ms for display on. Cycle time includes sweep

correction off. Add 21 ms for display on. Cycle

and re-trace time.

time includes sweep and re-trace time.

Cycle time (ms) 1,2


Number of points

201

401

1601

16,001

Start 28 GHz, stop 30 GHz, IFBW = 35 kHz

Uncorrected and one-port cal

12

19

55

503

Two-port

cal

29

44

124

1112

Start 10 MHz, stop 10 GHz, IFBW = 35 kHz
Uncorrected and one-port cal

86

93

121

583

Two-port

cal

179

199

267

1301

Start 10 MHz, stop 20 GHz, IFBW = 35 kHz
Uncorrected and one-port cal

126

130

153

597

Two-port

cal

264

275

335

1321

Start 10 MHz, stop 40 GHz, IFBW = 35 kHz
Uncorrected and one-port cal

185

190

213

621

Two-port

cal

382

401

459

1374

Start 10 MHz, stop 50 GHz, IFBW = 35 kHz
Uncorrected and one-port cal

210

216

243

643

Two-port

cal

436

450

522

1405

Start 10 MHz, stop 67 GHz, IFBW = 35 kHz
Uncorrected

244

254

300

645

Corrected

502

524

591

1423

IF bandwidth (Hz)

Cycle time (ms) Cycle time (ms)

Option 080 enabled

40,000 11 100

35,000

12

101

30,000

13

102

20,000

16

106

10,000

30

127

7,000

38

138

5,000

50

152

3,000

74

182

1,000

274

326

300

694

782

100

1905

2054

30

6091

6355

10

17916

18372

Number of points Cycle time (ms)

3 6

11

6

51

7

101

9

201

12

401

18

801

30

1601

55

16,001

497

1. Typical performance.
2. Includes sweep time, retrace time and band-crossing time. Analyzer display

turned off with DISPLAY:ENABLE OFF. Add 21 ms for display on.

Data for one trace (S11) measurement.