E8361c – Atec Agilent-E8364C User Manual
Page 29
29
E8361C
Corrected system performance with 2.4 mm connectors
Description Specification
(dB)
10 to 200 MHz
1
200 MHz to 2 GHz
2 to 20 GHz
20 to 40 GHz
40 to 50 GHz
Directivity
32 55 49 43 41
Source
match
25 46 42 35 30
Load
match
24 43 41 37 36
Reflection tracking
±0.050 (+0.02/°C)
±0.030 (+0.02/°C)
±0.040 (+0.02/°C)
±0.060 (+0.02/°C)
±0.080 (+0.03/°C)
Transmission tracking
±0.100 (+0.02/°C)
±0.059 (+0.02/°C)
±0.079 (+0.02/°C)
±0.110 (+0.02/°C)
±0.125 (+0.03/°C)
Transmission uncertainty (specifications)
Reflection uncertainty (specifications)
Magnitude
0.01
0.1
1
10
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Transmission coefficient (dB)
E8361C full two port cal using N4693A
200 MHz to 2 GHz
2 to 20 GHz
20 to 40 GHz
40 to 50 GHz
S11 = S22 = 0
Source power = -15 dBm
Uncertainty (dB)
Phase
0.1
1
10
100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Transmission coefficient (dB)
E8361C full two port cal using N4693A
200 MHz to 2 GHz
2 to 20 GHz
20 to 40 GHz
40 to 50 GHz
S11 = S22 = 0
Source power = -15 dBm
Uncertainty (degrees)
Magnitude
0
0.01
0.02
0.03
0.04
0.05
0
0.2
0.4
0.6
0.8
1
Reflection coefficient (linear)
E8361C with N4693A
200 MHz to 2 GHz
2 to 20 GHz
20 to 40 GHz
40 to 50 GHz
S21 = S12 = 0
Source power = -15 dBm
Uncertainty (linear)
Phase
0
2
4
6
8
10
0
0.2
0.4
0.6
0.8
1
Reflection coefficient (linear)
E8361C with N4693A
200 MHz to 2 GHz
2 to 20 GHz
20 to 40 GHz
40 to 50 GHz
S21 = S12 = 0
Source power = -15 dBm
Uncertainty (degrees)
Standard configuration and standard power range (E8361C)
Applies to E8361C PNA Series analyzer, N4693A
(2.4 mm) ECal electronic calibration module, 85133F
flexible test port cable set, and a full two-port
calibration. (Specifications apply over environmental
temperature of 23 °C ±3 °C, with less than 1 °C
deviation from calibration temperature.)
1. Typical performance.