beautypg.com

E8361c – Atec Agilent-E8364C User Manual

Page 29

background image

29

E8361C

Corrected system performance with 2.4 mm connectors

Description Specification

(dB)

10 to 200 MHz

1

200 MHz to 2 GHz

2 to 20 GHz

20 to 40 GHz

40 to 50 GHz

Directivity

32 55 49 43 41

Source

match

25 46 42 35 30

Load

match

24 43 41 37 36

Reflection tracking

±0.050 (+0.02/°C)

±0.030 (+0.02/°C)

±0.040 (+0.02/°C)

±0.060 (+0.02/°C)

±0.080 (+0.03/°C)

Transmission tracking

±0.100 (+0.02/°C)

±0.059 (+0.02/°C)

±0.079 (+0.02/°C)

±0.110 (+0.02/°C)

±0.125 (+0.03/°C)

Transmission uncertainty (specifications)

Reflection uncertainty (specifications)

Magnitude

0.01

0.1

1

10

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Transmission coefficient (dB)

E8361C full two port cal using N4693A

200 MHz to 2 GHz
2 to 20 GHz
20 to 40 GHz
40 to 50 GHz

S11 = S22 = 0
Source power = -15 dBm

Uncertainty (dB)

Phase

0.1

1

10

100

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Transmission coefficient (dB)

E8361C full two port cal using N4693A

200 MHz to 2 GHz
2 to 20 GHz

20 to 40 GHz

40 to 50 GHz

S11 = S22 = 0
Source power = -15 dBm

Uncertainty (degrees)

Magnitude

0

0.01

0.02

0.03

0.04

0.05

0

0.2

0.4

0.6

0.8

1

Reflection coefficient (linear)

E8361C with N4693A

200 MHz to 2 GHz
2 to 20 GHz
20 to 40 GHz
40 to 50 GHz

S21 = S12 = 0
Source power = -15 dBm

Uncertainty (linear)

Phase

0

2

4

6

8

10

0

0.2

0.4

0.6

0.8

1

Reflection coefficient (linear)

E8361C with N4693A

200 MHz to 2 GHz
2 to 20 GHz
20 to 40 GHz

40 to 50 GHz

S21 = S12 = 0
Source power = -15 dBm

Uncertainty (degrees)

Standard configuration and standard power range (E8361C)
Applies to E8361C PNA Series analyzer, N4693A
(2.4 mm) ECal electronic calibration module, 85133F
flexible test port cable set, and a full two-port
calibration. (Specifications apply over environmental
temperature of 23 °C ±3 °C, with less than 1 °C
deviation from calibration temperature.)

1. Typical performance.