E8361c, Test port input – Atec Agilent-E8364C User Manual
Page 41
41
1. Stability is defined as a ratio measurement measured at the test port.
E8361C
Test port input
continued
Description Specification
Supplemental
Standard
Option
014
information
Reference level magnitude
Range
±500
dB
±500
dB
Resolution
0.001
dB
0.001
dB
Reference level phase
Range
±500°
±500°
Resolution
0.01°
0.01°
Stability magnitude
1
Typical ratio measurement:
Measured at the test port
10 to 45 MHz
±0.05 dB/°C
45 MHz to 20 GHz
±0.02 dB/°C
20 to 40 GHz
±0.02 dB/°C
40 to 50 GHz
±0.02 dB/°C
50 to 70 GHz
±0.04 dB/°C
Stability phase
1
Typical ratio measurement:
Measured at the test port
10 to 45 MHz
±0.5°/°C
45 MHz to 20 GHz
±0.2°/°C
20 to 40 GHz
±0.5°/°C
40 to 50 GHz
±0.8°/°C
50 to 70 GHz
±0.8°/°C
Damage input level
Test port 1 and 2
+27 dBm or ±40 VDC, typical
R1, R2 in
+15 dBm or ±15 VDC, typical
A, B in
+15 dBm or ±7 VDC, typical
Coupler thru (Option 014)
+27 dBm or ±40 VDC, typical
Coupler arm (Option 014)
+30 dBm or ±7 VDC, typical
Source out (reference)
+15 dBm or ±15 VDC, typical
Source out (test ports)
+27 dBm or ±5 VDC, typical