Atec Agilent-E8364C User Manual
Page 7
![background image](/manuals/318351/7/background.png)
7
E8362/3/4C
Corrected system performance with 2.4 mm connectors
Description Specification
(dB)
45 MHz to 2 GHz
2 to 20 GHz
20 to 40 GHz
40 to 50 GHz
Directivity 42 42 38 36
Source match
41
38
33
31
Load match
42
42
37
35
Reflection tracking
±0.001 (+0.02/°C)
±0.008 (+0.02/°C)
±0.020 (+0.02/°C)
±0.027 (+0.03/°C)
Transmission tracking
±0.010 (+0.02/°C)
±0.049 (+0.02/°C)
±0.105 (+0.02/°C)
±0.170 (+0.03/°C)
Transmission uncertainty (specifications)
Reflection uncertainty (specifications)
Magnitude
0.01
0.1
1
10
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Transmission coefficient (dB)
Uncertainty (dB)
E8363/4C full two port cal using 85056A
S11 = S22 = 0
Source power = -12 dBm
45 MHz to 2 GHz
2 to 20 GHz
20 to 40 GHz
40 to 50 GHz
Phase
0.1
1
10
100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Transmission coefficient (dB)
Uncertainty (degrees)
E8363/4C full two port cal using 85056A
S11 = S22 = 0
Source power = -12 dBm
45 MHz to 2 GHz
2 to 20 GHz
20 to 40 GHz
40 to 50 GHz
Magnitude
0
0.01
0.02
0.03
0.04
0.05
0
0.2
0.4
0.6
0.8
1
Reflection coefficient (linear)
Uncertainty (linear)
E8363/4C with 85056A
S21 = S12 = 0
Source power = -12 dBm
45 MHz to 2 GHz
2 to 20 GHz
20 to 40 GHz
40 to 50 GHz
Phase
0
2
4
6
8
10
0
0.2
0.4
0.6
0.8
1
Reflection coefficient (linear)
Uncertainty (deg)
E8363/4C with 85056A
S21 = S12 = 0
Source power = -12 dBm
45 MHz to 2 GHz
2 to 20 GHz
20 to 40 GHz
40 to 50 GHz
Standard configuration and standard power range (E8363/4C)
Applies to E8363/4C PNA Series analyzer,
85056A (2.4 mm) calibration kit, 85133F flexible
test
port cable set, and a full two-port calibration.
(Specifications apply over environmental temperature
of 23 °C ±3 °C, with less than 1 °C deviation from
calibration temperature.)