E8361C

Corrected system performance with 2.4 mm connectors

continued

Description Specification

(dB)

45 MHz to 2 GHz

2 to 20 GHz

20 to 40 GHz

40 to 50 GHz

Directivity 42 42 38 36
Source match

41

38

33

31

Load match

42

42

37

35

Reflection tracking

±0.001 (+0.02/°C)

±0.008 (+0.02/°C)

±0.020 (+0.02/°C)

±0.027 (+0.03/°C)

Transmission tracking

±0.035 (+0.02/°C)

±0.060 (+0.02/°C)

±0.146 (+0.02/°C)

±0.181 (+0.03/°C)

Transmission uncertainty (specifications)

Reflection uncertainty (specifications)

Magnitude

0.01

0.1

1

10

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Transmission coefficient (dB)

E8361C full two port cal using 85056A

45 MHz to 2 GHz
2 to 20 GHz
20 to 40 GHz
40 to 50 GHz

S11 = S22 = 0
Source power = -15 dBm

Uncertainty (dB)

Phase

0.1

1

10

100

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Transmission coefficient (dB)

E8361C full two port cal using 85056A

45 MHz to 2 GHz
2 to 20 GHz
20 to 40 GHz

40 to 50 GHz

S11 = S22 = 0
Source power = -15 dBm

Uncertainty (degrees)

Magnitude

0

0.01

0.02

0.03

0.04

0.05

0

0.2

0.4

0.6

0.8

1

Reflection coefficient (linear)

E8361C with 85056A

45 MHz to 2 GHz
2 to 20 GHz
20 to 40 GHz
40 to 50 GHz

S21 = S12 = 0
Source power = -15 dBm

Uncertainty (linear)

Phase

0

2

4

6

8

10

0

0.2

0.4

0.6

0.8

1

Reflection coefficient (linear)

E8361C with 85056A

45 MHz to 2 GHz

2 to 20 GHz

20 to 40 GHz
40 to 50 GHz

S21 = S12 = 0
Source power = -15 dBm

Uncertainty (degrees)

Standard configuration and standard power range (E8361C)
Applies to E8361C PNA Series analyzer, 85056A
(2.4 mm) calibration kit, 85133F flexible test
port cable set, and a full two-port calibration.
(Specifications apply over environmental tempera-
ture of 23 °C ±3 °C, with less than 1 °C deviation