E8361c – Atec Agilent-E8364C User Manual

Page 22

background image

22

E8361C

Corrected system performance with 1.85 mm connectors

continued

Description Specification

(dB)

30 to 40 GHz

40 to 50 GHz

50 to 60 GHz

60 to 67 GHz

Directivity 44 42 41 38
Source match

34

33

30

27

Load match

33

32

29

26

Reflection tracking

±0.060 (+0.02/°C)

±0.070 (+0.02/°C)

±0.080 (+0.02/°C)

±0.090 (+0.03/°C)

Transmission tracking

±0.087 (+0.02/°C)

±0.102 (+0.02/°C)

±0.121 (+0.02/°C)

±0.147 (+0.03/°C)

Transmission uncertainty (specifications)

Reflection uncertainty (specifications)

Magnitude

0.01

0.1

1

10

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Transmission coefficient (dB)

E8361C full two port cal using N4694A

30 to 40 GHz

40 to 50 GHz

50 to 60 GHz

60 to 67 GHz

S11 = S22 = 0
Source power = -15 dBm

Uncertainty (linear)

Uncertainty (linear)

Uncertainty (linear)Uncertainty (linear)Uncertainty (linear)

Uncertainty (dB)U

n

ce

rt

ai

n

ty

(

dB

)

Phase

0.1

1

10

100

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Transmission coefficient (dB)

E8361C full two port cal using N4694A

30 to 40 GHz

40 to 50 GHz

50 to 60 GHz

60 to 67 GHz

S11 = S22 = 0
Source power = -15 dBm

Uncertainty (degrees)

Magnitude

0

0.01

0.02

0.03

0.04

0.05

0

0.2

0.4

0.6

0.8

1

Reflection coefficient (linear)

E8361C with N4694A

30 to 40 GHz

40 to 50 GHz

50 to 60 GHz
60 to 67 GHz

S21 = S12 = 0
Source power = -15 dBm

Uncertainty (linear)

Phase

0

2

4

6

8

10

0

0.2

0.4

0.6

0.8

1

Reflection coefficient (linear)

E8361C with N4694A

30 to 40 GHz

40 to 50 GHz

50 to 60 GHz

60 to 67 GHz

S21 = S12 = 0
Source power = -15 dBm

Uncertainty (deg)

Standard configuration and standard power range (E8361C)
Applies to E8361C PNA Series analyzer, N4694A
(1.85 mm) ECal electronic calibration module, N4697E/F
flexible test port cable set, and a full two-port calibration.
(Specifications apply over environmental temperature
of 23 °C ±3 °C, with less than 1 °C deviation from
calibration temperature.)