E8361c, Test port input – Atec Agilent-E8364C User Manual
Page 39
![background image](/manuals/318351/39/background.png)
39
E8361C
Test port input
continued
Description
Specification
Option 014 or
Supplemental
Standard
Option 014 and UNL (typ.)
information
Direct receiver access
input noise floor
1
10 Hz IF bandwidth
10 to 45 MHz
2
< -106 dBm
45 to 500 MHz
4, 5
< -105 dBm
Online Help also includes the
500 MHz to 2 GHz
< -125.5 dBm
category “Direct receiver access
2 to 10 GHz
< -125 dBm
input noise floor, Option 080 enabled”
10 to 24 GHz
< -128 dBm
24 to 30 GHz
< -117.5 dBm
30 to 40 GHz
< -117 dBm
Option 016 degrades
40 to 45 GHz
< -115 dBm
performance by 2 dB.
45 to 50 GHz
< -112.5 dBm
50 to 60 GHz
< -111 dBm
60 to 67 GHz
< -108 dBm
Option 016 degrades
67 to 70 GHz
2
< -107 dBm
performance by 3 dB.
1 kHz IF bandwidth
10 to 45 MHz
2
< -86 dBm
45 to 500 MHz
4, 5
< -85 dBm
500 MHz to 2 GHz
< -105.5 dBm
2 to 10 GHz
< -105 dBm
10 to 24 GHz
< -108 dBm
24 to 30 GHz
< -97.5 dBm
30 to 40 GHz
< -97 dBm
Option 016 degrades
40 to 45 GHz
< -95 dBm
performance by 2 dB.
45 to 50 GHz
< -92.5 dBm
50 to 60 GHz
< -91 dBm
60 to 67 GHz
< -88 dBm
Option 016 degrades
67 to 70 GHz
2
< -87 dBm
performance by 3 dB.
Description Specification
Supplemental
Receiver compression level (measured at Test Ports)
Supplemental
Specifications
Information
Standard
Option 014
Option 014 and UNL
Typical
10 to 45 MHz
2, 3
negligible
negligible
negligible
negligible
45 to 500 MHz
3, 6
<0.1 dB at -9.5 dBm
7
<0.1 dB at -9.5 dBm
7
<0.1 dB at -9.5 dBm
7
<0.1 dB at +0.5 dBm
7
and <0.25 dB at -3 dBm
and <0.25 dB at -3 dBm
and <0.25 dB at -3 dBm
and <0.25 dB at +8 dBm
500 MHz to 5 GHz
<0.1 dB at -8 dBm
7
<0.1 dB at -8 dBm
7
<0.1 dB at -7 dBm
7
<0.1 dB at -4 dBm
7
and <0.25 dB at -1 dBm
and <0.25 dB at -1 dBm
and <0.25 dB at 0 dBm
and <0.25 dB at +3 dBm
5 to 30 GHz
<0.1 dB at -8.5 dBm
7
<0.1 dB at -8.5 dBm
7
<0.1 dB at -6 dBm
7
<0.1 dB at -1 dBm
7
and <0.25 dB at -2 dBm
and <0.25 dB at -2 dBm
and <0.25 dB at +1 dBm
and <0.25 dB at +6 dBm
30 to 67 GHz
< 0.1 dB at -10.5 dBm
7
< 0.1 dB at -8 dBm
7
<0.1 dB at –9.5 dBm
7
<0.1 dB at -2 dBm
7, 8
and <0.15 dB at -7 dBm
and <0.15 dB at -4 dBm
and <0.15 dB at -6 dBm
and <0.15 dB at +2 dBm
8
67 to 70 GHz
2
<0.1 dB at -2 dBm
7, 8
and <0.15 dB at +2 dBm
8
1. Total average (rms) noise power calculated as mean value of a linear magnitude
trace expressed in dBm.
2. Typical performance.
3. Coupler roll-off will reduce compression to a negligible level below 500 MHz.
4. Noise floor may be degraded by 10 dB at particular frequencies
(multiples of 5 MHz) due to spurious receiver residuals.
5. Specified value is for worst-case noise floor at 45 MHz.
6. Specified value is for worst-case compression at 500 MHz.
7. This compression level comes from the dynamic accuracy curve with -30 dBm
reference test port power.
8. Option 016 degrades performance by 3 dB.