beautypg.com

E8361c, Test port input – Atec Agilent-E8364C User Manual

Page 39

background image

39

E8361C

Test port input

continued

Description

Specification

Option 014 or

Supplemental

Standard

Option 014 and UNL (typ.)

information

Direct receiver access
input noise floor

1

10 Hz IF bandwidth
10 to 45 MHz

2

< -106 dBm

45 to 500 MHz

4, 5

< -105 dBm

Online Help also includes the

500 MHz to 2 GHz

< -125.5 dBm

category “Direct receiver access

2 to 10 GHz

< -125 dBm

input noise floor, Option 080 enabled”

10 to 24 GHz

< -128 dBm

24 to 30 GHz

< -117.5 dBm

30 to 40 GHz

< -117 dBm

Option 016 degrades

40 to 45 GHz

< -115 dBm

performance by 2 dB.

45 to 50 GHz

< -112.5 dBm

50 to 60 GHz

< -111 dBm

60 to 67 GHz

< -108 dBm

Option 016 degrades

67 to 70 GHz

2

< -107 dBm

performance by 3 dB.

1 kHz IF bandwidth
10 to 45 MHz

2

< -86 dBm

45 to 500 MHz

4, 5

< -85 dBm

500 MHz to 2 GHz

< -105.5 dBm

2 to 10 GHz

< -105 dBm

10 to 24 GHz

< -108 dBm

24 to 30 GHz

< -97.5 dBm

30 to 40 GHz

< -97 dBm

Option 016 degrades

40 to 45 GHz

< -95 dBm

performance by 2 dB.

45 to 50 GHz

< -92.5 dBm

50 to 60 GHz

< -91 dBm

60 to 67 GHz

< -88 dBm

Option 016 degrades

67 to 70 GHz

2

< -87 dBm

performance by 3 dB.

Description Specification

Supplemental

Receiver compression level (measured at Test Ports)

Supplemental

Specifications

Information

Standard

Option 014

Option 014 and UNL

Typical

10 to 45 MHz

2, 3

negligible

negligible

negligible

negligible

45 to 500 MHz

3, 6

<0.1 dB at -9.5 dBm

7

<0.1 dB at -9.5 dBm

7

<0.1 dB at -9.5 dBm

7

<0.1 dB at +0.5 dBm

7

and <0.25 dB at -3 dBm

and <0.25 dB at -3 dBm

and <0.25 dB at -3 dBm

and <0.25 dB at +8 dBm

500 MHz to 5 GHz

<0.1 dB at -8 dBm

7

<0.1 dB at -8 dBm

7

<0.1 dB at -7 dBm

7

<0.1 dB at -4 dBm

7

and <0.25 dB at -1 dBm

and <0.25 dB at -1 dBm

and <0.25 dB at 0 dBm

and <0.25 dB at +3 dBm

5 to 30 GHz

<0.1 dB at -8.5 dBm

7

<0.1 dB at -8.5 dBm

7

<0.1 dB at -6 dBm

7

<0.1 dB at -1 dBm

7

and <0.25 dB at -2 dBm

and <0.25 dB at -2 dBm

and <0.25 dB at +1 dBm

and <0.25 dB at +6 dBm

30 to 67 GHz

< 0.1 dB at -10.5 dBm

7

< 0.1 dB at -8 dBm

7

<0.1 dB at –9.5 dBm

7

<0.1 dB at -2 dBm

7, 8

and <0.15 dB at -7 dBm

and <0.15 dB at -4 dBm

and <0.15 dB at -6 dBm

and <0.15 dB at +2 dBm

8

67 to 70 GHz

2

<0.1 dB at -2 dBm

7, 8

and <0.15 dB at +2 dBm

8

1. Total average (rms) noise power calculated as mean value of a linear magnitude

trace expressed in dBm.

2. Typical performance.
3. Coupler roll-off will reduce compression to a negligible level below 500 MHz.
4. Noise floor may be degraded by 10 dB at particular frequencies

(multiples of 5 MHz) due to spurious receiver residuals.

5. Specified value is for worst-case noise floor at 45 MHz.
6. Specified value is for worst-case compression at 500 MHz.
7. This compression level comes from the dynamic accuracy curve with -30 dBm

reference test port power.

8. Option 016 degrades performance by 3 dB.