beautypg.com

E8361c, Test port input – Atec Agilent-E8364C User Manual

Page 40

background image

40

E8361C

Test port input

continued

Description Specification

Option

UNL

Supplemental

Standard or Option 014

and Option 014

information

Third Order Intercept

1

– Tone spacing from 100 kHz to 5 MHz

Typical:

10 to 500 MHz

+30 dBm

500 MHz to 24 GHz

+24 dBm

24 to 40 GHz

+23 dBm

40 to 50 GHz

+24 dBm

50 to 67 GHz

+26 dBm

Third Order Intercept

1

– Tone spacing from 5 to 20 MHz

Typical:

10 to 500 MHz

Not applicable

500 MHz to 24 GHz

+20 dBm

24 to 40 GHz

+20 dBm

40 to 50 GHz

+22 dBm

50 to 67 GHz

+24 dBm

Third Order Intercept

1

– Tone spacing from 20 to 50 MHz

Typical:

10 to 500 MHz

Not applicable

500 MHz to 24 GHz

+26 dBm

24 to 40 GHz

+24 dBm

40 to 50 GHz

+25 dBm

50 to 67 GHz

+27 dBm

System compression level – at maximum leveled output power

See Dynamic Accuracy Chart
Trace noise magnitude
10 to 45 MHz

2

< 0.150 dB rms

< 0.150 dB rms

45 to 500 MHz

4, 5

< 0.010 dB rms

< 0.010 dB rms

1 kHz IF bandwidth

500 MHz to 24 GHz

< 0.006 dB rms

< 0.006 dB rms

ratio measurement, nominal

24 to 67 GHz

< 0.006 dB rms

< 0.009 dB rms

power at test port

67 to 70 GHz

2

< 0.006 dB rms

< 0.009 dB rms

Trace noise magnitude

2

– Option 080 enabled

3

10 to 45 MHz

< 0.150 dB rms

< 0.150 dB rms

45 to 500 MHz

4, 5

< 0.010 dB rms

< 0.010 dB rms

1 kHz IF bandwidth

500 MHz to 24 GHz

< 0.006 dB rms

< 0.006 dB rms

ratio measurement, nominal

24 to 67 GHz

< 0.009 dB rms

< 0.012 dB rms

power at test port

67 to 70 GHz

< 0.009 dB rms

< 0.012 dB rms

Trace noise phase
10 to 45 MHz

2

< 0.800° rms

< 0.800° rms

45 to 500 MHz

5

< 0.100° rms

< 0.100° rms

1 kHz IF bandwidth

500 MHz to 24 GHz

< 0.060° rms

< 0.060° rms

ratio measurement, nominal

24 to 67 GHz

< 0.100° rms

< 0.100° rms

power at test port

67 to 70 GHz

2

< 0.100° rms

< 0.100° rms

Trace noise phase

2

– Option 080 enabled

3

10 to 45 MHz

< 0.800° rms

< 0.800° rms

1 kHz IF bandwidth

45 to 500 MHz

5

< 0.100° rms

< 0.100° rms

ratio measurement, nominal

500 MHz to 24 GHz

< 0.060° rms

< 0.060° rms

power at test port

24 to 67 GHz

< 0.100° rms

< 0.100° rms

67 to 70 GHz

< 0.100° rms

< 0.100° rms

1. Third order intercept is a typical specification that applies while the network

analyzer receiver is in its linear range.

2. Typical performance.
3. 0 Hz offset.
4. Trace noise magnitude may be degraded to 20 mdB rms at harmonic frequencies

of the first IF (8.33 MHz) below 80 MHz.

5. Specified value is for worst-case noise floor at 45 MHz.