Uncorrected system performance 1 – Atec Agilent-E8364C User Manual
Page 11
![background image](/manuals/318351/11/background.png)
11
E8362/3/4C
Uncorrected system performance 1
Description Specification
Supplemental
information
Directivity
Typical:
10 to 45 MHz
2
23 dB
23 dB
45 MHz to 2 GHz
24 dB
29 dB
2 to 10 GHz
22 dB
25 dB
10 to 20 GHz
16 dB
20 dB
20 to 40 GHz
16 dB
20 dB
40 to 45 GHz
15 dB
18 dB
45 to 50 GHz
13 dB
18 dB
Source match - standard
Typical:
10 to 45 MHz
2
11 dB
12 dB
45 MHz to 2 GHz
23 dB
27 dB
2 to 10 GHz
16 dB
19 dB
10 to 20 GHz
14 dB
19 dB
20 to 40 GHz
10 dB
14 dB
40 to 45 GHz
9 dB
13.5 dB
45 to 50 GHz
7.5 dB
10 dB
Source match - Option UNL, 014, or UNL and 014
Typical:
10 to 45 MHz
2
11 dB
12 dB
45 MHz to 2 GHz
18 dB
22.5 dB
2 to 10 GHz
14 dB
18 dB
10 to 20 GHz
12 dB
15 dB
20 to 40 GHz
9 dB
11 dB
40 to 45 GHz
8 dB
13 dB
45 to 50 GHz
6 dB
9 dB
Load match - standard
Typical:
10 to 45 MHz
2
11 dB
12 dB
45 MHz to 2 GHz
23 dB
29 dB
2 to 10 GHz
14 dB
16 dB
10 to 20 GHz
10 dB
12 dB
20 GHz to 40 GHz
9 dB
12 dB
40 to 45 GHz
9 dB
13 dB
45 to 50 GHz
8 dB
10 dB
Load match - Option UNL, 014, or UNL and 014
Typical:
10 to 45 MHz
2
11 dB
12 dB
45 MHz to 2 GHz
17 dB
21.5 dB
2 to 10 GHz
13 dB
16.5 dB
10 to 20 GHz
10 dB
13 dB
20 to 40 GHz
9 dB
11 dB
40 to 45 GHz
9 dB
13 dB
45 to 50 GHz
7 dB
9.5 dB
Reflection tracking
Typical:
10 to 45 MHz
2
±1.5 dB
45 MHz to 20 GHz
±1.5 dB
20 to 40 GHz
±1.5 dB
40 to 50 GHz
±2.0 dB
Transmission tracking
3
Typical:
10 to 45 MHz
2
±3.0 dB
45 MHz to 2 GHz
±1.5 dB
2 to 10 GHz
±2.0 dB
10 to 20 GHz
±2.5 dB
20 to 40 GHz
±3.5 dB
40 to 45 GHz
±4.0 dB
45 to 50 GHz
±4.5 dB
1. Specifications apply over environment temperature of 23 °C ±3 °C, with less than 1 °C
deviation from the calibration temperature.
2. Typical performance.
3. Transmission tracking performance is strongly dependent on cable used. These
typical specifications are based on the use of an Agilent through cable, part
number
85133-60016.