E8361C
Corrected system performance with 1.85 mm connectors
continued
Description Specification
(dB)
10 to 45 MHz
1
45 MHz to 2 GHz
2 to 10 GHz
10 to 20 GHz
Directivity 35 35 41 38
Source match
34
34
44
40
Load match
34
35
41
37
Reflection tracking
±0.019 (+0.02/°C)
±0.019 (+0.02/°C)
±0.010 (+0.02/°C)
±0.033 (+0.02/°C)
Transmission tracking
±0.177 (+0.02/°C)
±0.093 (+0.02/°C)
±0.053 (+0.02/°C)
±0.096 (+0.02/°C)
Transmission uncertainty (specifications)
Reflection uncertainty (specifications)
Magnitude
0.01
0.1
1
10
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Transmission coefficient (dB)
E8361C fully optioned
2
full two port cal using 85058B (Exp math)
10 to 45 MHz
1
45 MHz to 2 GHz
2 to 10 GHz
10 to 20 GHz
S11 = S22 = 0
Source power = -20 dBm
Uncertainty (dB)
Phase
0.1
1
10
100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Transmission coefficient (dB)
E8361C fully optioned
2
full two port cal using 85058B (Exp math)
10 to 45 MHz
1
45 MHz to 2 GHz
2 to 10 GHz
10 to 20 GHz
S11 = S22 = 0
Source power = -20 dBm
Uncertainty (degrees)
Magnitude
0
0.01
0.02
0.03
0.04
0.05
0
0.2
0.4
0.6
0.8
1
Reflection coefficient (linear)
E8361C fully optioned
2
with 85058B (Exp math)
10 to 45 MHz
1
45 MHz to 2 GHz
2 o 10 GHz
10 to 20 GHz
S21 = S12 = 0
Source power = -20 dBm
Uncertainty (linear)
Phase
0
2
4
6
8
10
0
0.2
0.4
0.6
0.8
1
Reflection coefficient (linear)
E8361C fully optioned
2
with 85058B (Exp math)
10 to 45 MHz
1
45 MHz to 2 GHz
2 to 10 GHz
10 to 20 GHz
S21 = S12 = 0
Source power = -20 dBm
Uncertainty (degrees)
Fully optioned (E8361C with options 014/
UNL/080/081/016)
Applies to E8361C PNA Series analyzer,
85058B (1.85 mm) calibration kit, N4697F flexible
test port cable set, and a full two-port calibration.
(Specifications apply over environmental tempera-
ture of 23 °C ±3 °C, with less than 1 °C deviation
from calibration temperature.)
1. Typical performance.
2. Configurable Test Set, Extended Power Range and Bias-Tees, Receiver Attenuators, Frequency Offset Mode, and