beautypg.com

Test port output – Atec Agilent-E8364C User Manual

Page 13

background image

13

E8362/3/4C

Test port output

Description Specification

Supplemental

information

Standard

014

UNL

UNL

and

014

Frequency range
E8362C

10 MHz to 20 GHz

E8363C

10 MHz to 40 GHz

E8364C

10 MHz to 50 GHz

Nominal power

2

E8362C

0 dBm

-5 dBm

-5 dBm

-5 dBm

E8363/4C

-12 dBm

-17 dBm

-17 dBm

-17 dBm

Frequency resolution 1 Hz

1 Hz

1 Hz

1 Hz

CW accuracy

±

1ppm ±

1ppm ±

1ppm ±

1ppm

Frequency stability

±1 ppm, 0 to 40 °C, typical

±0.2

ppm/yr,

typical

Power level accuracy

1

10 to 45 MHz

3

±2.0 dB

±2.0 dB

±2.0 dB

±2.0 dB

45 MHz to 10 GHz ±1.5 dB

±1.5 dB

±1.5 dB

±1.5 dB

Variation from nominal

10 to 20 GHz

±2.0 dB

±2.0 dB

±2.0 dB

±2.0 dB

power in range 0

20 to 40 GHz

±3.0 dB

±3.0 dB

±3.0 dB

±3.0 dB

(step attenuator at 0 dB).

40 to 45 GHz

±3.0 dB

±3.5 dB

±3.0 dB

±3.5 dB

45 to 50 GHz

±3.0 dB

±4.0 dB

±3.0 dB

±4.0 dB

Power level linearity

6

10 to 45 MHz

3

±1.0 dB

4

±1.0

dB

4

±1.0

dB

4

±1.0

dB

4

45 MHz to 20 GHz ±1.0 dB

4

±1.0

dB

4

±1.0

dB

4

±1.0

dB

4

Test reference is at the

20 to 40 GHz

±1.0 dB

4

±1.0

dB

4

±1.0

dB

4

±1.0

dB

4

nominal power level

40 to 50 GHz

±1.0 dB

4

±1.0

dB

4

±1.0

dB

4

±1.0

dB

4

(step attenuator at 0 dB).

Power range

1, 5, 7

10 to 45 MHz

3

-25 to +2 dB

-25 to +2 dBm

-87 to +2 dBm

-87 to +2 dBm

45 MHz to 10 GHz -25 to +5 dB

-25 to +5 dBm

-87 to +3 dBm

-87 to +3 dBm

8

10 to 20 GHz

-24 to +3 dB

-25 to +2 dBm

-86 to +1 dBm

-87 to 0 dBm

9

20 to 30 GHz

-23 to 0 dBm

-25 to -2 dBm

-85 to -2 dBm

-87 to -4 dBm

30 to 40 GHz

-23 to -4 dBm

-25 to - 6 dBm

-85 to -6 dBm

-87 to -8 dBm

40 to 45 GHz

-25 to -5 dBm

-27 to -7 dBm

-87 to -9 dBm

-87 to -11 dBm

45 to 50 GHz

-25 to -10 dBm

-27 to -12 dBm

-87 to -15 dBm

-87 to -17 dBm

Power sweep range (ALC)
10 to 45 MHz

3

27 dB

27 dB

29 dB

29 dB

45 MHz to 10 GHz 30 dB

30 dB

30 dB

30 dB

10

ALC range starts at

10 to 20 GHz

27 dB

27 dB

27 dB

27 dB

11

maximum leveled output

20 to 30 GHz

23 dB

23 dB

23 dB

23 dB

power and decreases by

30 to 40 GHz

19 dB

19 dB

19 dB

19 dB

power level indicated in

40 to 45 GHz

20 dB

20 dB

18 dB

16 dB

the table.

45 to 50 GHz

15 dB

15 dB

12 dB

10 dB

Power resolution

0.01 dB

0.01 dB

0.01 dB

0.01 dB

1. Test port output is a specification when the source is set to port 1 and a

characteristic when the source is set to port 2.

2. Preset power.
3. Typical performance.
4. ±1.5 dB for power -23 dBm.
5. Power to which the source can be set and phase lock is assured.
6. Power level linearity is a specification when the source is set to port 1 and a

typical when the source is set to port 2.

7. Test port power is specified into nominal 50 ohms.
8. Option H11 decreases maximum power level by 1 dB.
9. Option H11 decreases maximum power level by 2 dB.
10. Option H11 decreases power level by 1 dB.
11. Option H11 decreases power level by 2 dB.