Electromagnetic Flow Meter
FW: 842018/842009
Remote Flow Sensor
Connect the remote flow sensor to the flow converter with the included
cable (part number 691080).
Important: Do not use other cable types!
Important: Terminals 1 and 4 must be shunted!
The extension cable length between converter and sensor must under
normal circumstances not exceed 50 meters. Power lines running in paral-
lel and noisy environments may reduce the max. length to less than 50
meters. Consult MJK personnel for advice and guidance.
Remote Flow Sensor
691080
Terminal
Signal
Designation
Colour
before March 2011
Colour
from March 2011
1
Liquid GND
Shield
-
-
2
E1
Wire pair no. 1
Red
Red
3
E2
Wire pair no. 1
White
White
4
GND
Wire pair no. 3
Black
Black
5
L1
Wire pair no. 2
Green
Blue
6
L2
Wire pair no. 2
White
Brown
7, 8 and 9
not used
Converter Connection Board - Local
Red
White
Blac
k
Blue
Bro
wn
F
or attached
Displa
y Unit
F
or remote
Displa
y Unit
Digital output 1
DO
1
D
I
1
DO
2
Digital output
2
Digital input
10-30
V DC
4-20 mA output
P
ower supply
Electronic rela
y
max.
50
V DC 120 mA
Electromechanical rela
y
max.
50
V
A
C
/ 1
A
4-20 mA max.
800
Ω
Fuse rating:
125 mA
T @ 115
V
A
C
Connection for
Flow Sensor
63 mA
T @ 230
V
A
C