3 distributed mounting of miq modules, Distributed mounting of miq modules -21, Section 3.5.3 – YSI IQ SensorNet 2020 XT Terminal User Manual
Page 51
System 2020 XT USB
Installation
3 - 21
ba76019e01
12/2012
3.5.3
Distributed mounting of MIQ modules
General information
The following IQ S
ENSOR
N
ET
cables can be used for distributed
mounting:
SNCIQ cable
SNCIQ/UG earth cable - suitable for underground laying according
to VDE 01816, Part 2 and DIN/VDE 0891, Part 6.
The cables are delivered as piece goods (please specify length when
ordering!).
Each MIQ module has a row of terminal strips in the housing. The two
3-fold terminal strips on the right are always reserved for the
connection of further IQ S
ENSOR
N
ET
components (designation
"SENSORNET 1" and "SENSORNET 2").
Some MIQ modules have additional "SENSORNET" connections
("SENSORNET 3" and "SENSORNET 4") on the left of the terminal
strip.
Caution
The IQ S
ENSOR
N
ET
cable may only be connected to the
SENSORNET connections. No wire of the cable may be connected
with an external electrical potential. Otherwise, malfunctions
could occur.
Note
IQ S
ENSOR
N
ET
cables must always be installed separately at a
minimum distance of 20 cm from other lines that carry a voltage greater
than 60 V.
Materials required
1 x SNCIQ or SNCIQ/UG connection cable (see chapter 10
Wire end sleeves for 0.75 mm
2
wire cross-section with matching
crimping tool
1 x cable gland with seal (scope of delivery of MIQ module).
Tools
Cable stripping knife
Wire stripper
Phillips screw driver
Small screw driver.
9
Close the back MIQ module and fix it with the two countersunk
screws (pos. 5 in Fig. 3-13).