3 availability of the system, Availability of the system -19 – YSI IQ SensorNet 2020 XT Terminal User Manual
Page 25
System 2020 XT USB
Overview
1 - 19
ba76019e01
12/2012
1.4.3
Availability of the system
As a basic principle, the IQ S
ENSOR
N
ET
offers a very high availability.
This is achieved by
The two-wire connection technique that is not interference-prone
The digital signal transmission resulting from this
Tools for error diagnosis
– LEDs on the MIQ modules to check the available voltage
– Log book messages with instructions to eliminate problems
The possibility to simply assign substitute components for
components to be exchanged.
Replacing components is simple and can be performed very quickly
(see section 7.4). The system configuration (Settings of sensors and
diff. sensors and Settings of outputs and links) and the measured
data are stored in the controller and remain stored when other
components are used.
Controller
No current measured data can be displayed,
processed or stored.
If communication with the controller is disturbed,
output modules activate the error behavior after
2 minutes (see component operating manual of
the output module).
Terminal
The measured value display is not updated.
The display does not react to entries.
The system continues working.
Stored measured data and the system
configuration remain stored in the MIQ/
TC 2020 XT.
Power supply
module
like power failure (see section 1.4.1)
Component
Behavior