YSI IQ SensorNet 2020 XT Terminal User Manual
Page 173
System 2020 XT USB
What to do if ...
7 - 13
ba76019e01
12/2012
Operating sequence in
case 2
Case 1:
The serial number of the
MIQ output module is identical
with the serial number in an
inactive dataset.
The connected MIQ output
module is automatically
assigned to the inactive dataset
and starts to operate again.
Example: In the case of repair.
Case 2:
The type of the MIQ output
module is identical with the
type in an inactive dataset (or
several inactive datasets), but
the serial numbers differ.
Operator intervention is required
here. The connected MIQ output
module can:
– be assigned to the inactive
dataset (or one of the inactive
datasets).
Example: Replacing an
MIQ output module.
– be included as a new module
in the list of outputs.
The operating sequence to do
this is described below.
1
Remove the (defective) MIQ output module. The output
module is removed in the reverse order to the installation. The
installation is described in the I
NSTALLATION
chapter of the
respective component operating manual.
2
Install the new MIQ output module (I
NSTALLATION
chapter of
the respective component operating manual).
3
Change to the measured value display with
database is updated. The following display appears (example):