2 explanation of the message codes, 1 error messages, 2 explanation of the message codes -2 – YSI IQ SensorNet 2020 XT Terminal User Manual
Page 189: 1 error messages -2

System 2020 XT USB
lists
10 - 2
ba76019e01
12/2012
11.2
Explanation of the message codes
The log book contains a list with all the messages from all modules. Each message consists of message
code, date and time. You can obtain more detailed information by opening the full message text (see
section 4.5).
The full message text comes from the component that has triggered the message. Therefore, these
texts are only available from components which are connected to the system and ready for operation.
If a message text is not available, because the component is not connected to the system, you can look
up the texts of the messages in the operating manual of the respective component.
The following lists contain the message codes and the related message texts that are shown on the
display. Error messages and info messages are listed separately.
General explanations on the subjects of messages, message codes and log book are given in section
4.5 of this system operating manual.
11.2.1 Error messages
Message code
Message text
EAB141
Error at automatic air pressure measurement
An air pressure value of 1013 mbar is used for air pressure
compensation
* Component has been removed from the system, insert component
again
* Component defective, contact service
EI4141
Max. components of this component type exceeded
Danger of system overload
* Check and adapt the components for this type
EI5141
Component cannot be reached or not present
* Component has been removed from the system, insert component
again
* Connex. to component cut.,
Check sys. installation acc. to op. manual
EI6141
Incompatible terminal software
* Contact service
EI7141
Incompatible controller software
* Contact service