2 erasing a sensor-sensor link, Erasing a sensor-sensor link – YSI IQ SensorNet 2020 XT Terminal User Manual
Page 142

Settings/setup
System 2020 XT USB
5 - 22
ba76019e01
12/2012
5.7.2
Erasing a Sensor-sensor link
Note
The Sensor-sensor link function is automatically deactivated if the
measured parameter of the linked sensor is changed (e.g. by switching
the display of the TetraCon
®
700 IQ conductivity sensor from salinity to
conductivity).
1
Using , open the Settings menu.
2
Using <
><> and
System settings -> Sensor-sensor link.
The Sensor-sensor link overview opens.
The displayed list includes all sensors with which the
correction of an influence quantity is possible (e.g. FDO
®
700
IQ).
3
Using <
><>, highlight a sensor and confirm with
The displayed list includes all linked sensors.
4
Using <
><>, highlight a sensor and confirm with
The displayed list includes all sensors that measure an
influence quantity with the correct unit
(e.g. TetraCon
®
700 IQ, which measures the salinity in mg/L),
and the menu item, Erase link.
5
Using <
><>, highlight the menu item, Erase link, and
confirm with
The sensor-sensor link is deleted.
6
Leave the system settings with
In the measured value display, the measured value is
corrected by the manually entered value.