beautypg.com

Dismantling the dilution cross, Figure 13, Exploded drawing of dilution unit – M&C TechGroup SP2000H_DIL_BR Operator's manual User Manual

Page 27

background image

27

Gas sampling and gas conditioning technology

2-1.1.7-ME

16.2

DISMANTLING THE DILUTION CROSS

Figure 11 shows an exploded drawing of the dilution unit.

Nut 1

Nut 2

Nut 3

Nut 4

Nut 5

Nut 6

Critical orifice

Mark

Injector

Mark

Mark

Tube sample gas
out

Block

180°C Version

O-rings

Figure 13

Exploded drawing of dilution unit

The following step-by-step procedure is recommended:

Remove probe insulating cover (see 16.1).

Loosen knurled screw for fixing heat conducting plates.

Remove heat conducting plates.

Remove all pipe connections on dilution unit (crosspiece). These are:

- vacuum pressure gauge nut 6
- dilution gas inlet nut 4
- sample gas outlet nut 5 (remove sample gas out pipe at upper elbow union. Do not remove

pipe connected to crosspiece)

- nut 1.

W A R N I N G !


Do not remove nut 2 until the dilution cross is removed from the
probe or the orifice will be broken !


The dilution cross is now ready for further maintenance and repair work.