24

Gas sampling and gas conditioning technology

2-1.1.7-ME

16

MAINTENANCE

Prior to maintenance and repairs, system and process-specific safety measures must be observed.

W A R N I N G !


Aggressive condensation possible. Safety goggles and appropri-
ate protective clothing should be worn!

Caution: Do not touch the probe surface during use as this can
cause burns due to the high surface temperatures. Protective
gloves must be worn and the probe protected against unauthorised
access!

Before carrying out maintenance work on electrical components,
the mains voltage must be disconnected in all poles. This applies
similarly to any connected alarm and control circuits!


No recommendations for maintenance intervals can be given. Maintenance intervals must be deter-
mined depending on the particular process conditions and specific application.

Probe maintenance is limited mainly to replacement of the filter elements, inspection of the seals and
maintenance of the dilution systems.

N O T E !

The probe does not need to be removed for repairs or maintenance.
The probe will need to be removed for replacement of the process-
sided sample pipe.