Fault symptom codes, Fault symptom codes -28, Table 5-6. csl fscs -28 – StorageTek 9840 User Manual
Page 144
![background image](/manuals/135996/144/background.png)
Error and Recovery
5-28
Sixth Edition
95741
Fault Symptom Codes
Perform the following sequence of checks if a CSL error occurs.
1. Check the CSL
operator
indicator to see if it is flashing (operator summoned)
or on solid (failure). Clear the obvious error.
2. If the CSL does not indicate any problems, check the FSC list (
3. Check sense data at the host for any FSC in addition to the information
provided in
.
Note: Some malfunctions are caused by communications errors between the
CSL and the drive, and also by internal software. If the malfunctions
continue, force a dump of the drive for engineering evaluation.
provides the 4EXX FSCs, suspected malfunction, and possible fix.
Table 5-6. CSL FSCs
FSC
Malfunction
Fix
4E04
Cartridge installed incorrectly (180
degrees out, cartridge reels to left).
Drive replaced within the CSL, and
4E04 displays when cartridge is
loaded.
Insert cartridge correctly.
Install cartridge with reels to
the right. See
Drive may not be aligned
correctly within CSL. Have a
qualified technician remove,
replace, and align the drive.
4E05
SCSI load command rejected due to
no CSL being attached to the drive.
Check the black control cable
between the drive and CSL.
Final action: have a qualified
technician replace the drive.
4E0C
SCSI load command cannot be
performed. The CSL is not in the
System mode.
Select the System mode.
4E0D
SCSI load command cannot be
performed. There is no cartridge in
the input bin or feed channel.
Load cartridge.
Note: The 4E0D FSC may occur after a cleaning cartridge has been
unloaded.
4E11
ESCON load command is rejected
due to no CSL being attached to the
drive.
Have a qualified technician
check the black control cable
between the drive and CSL.
Final action: have a qualified
technician replace the drive.