- 49 -












Reference mark signal
Oscilloscope display x - t

Change the trigger setting of the oscilloscope basic setting
as follows:
Trigger channel A
Manual triggering (AC or DC)
– Trigger

negative edge

Traverse the reference mark to be checked
oscillatorily (“forward“/“backward“).

At the oscilloscope turn the trigger
potentiometer to set the trigger threshold
(LEVEL) such that the reference mark
signal is depicted as “stationary“ image
on the screen.

Note:
Schematic display of Ue1+2 on the oscilloscope.
The actual amplitude is higher than displayed.