beautypg.com

6 specifications of ttl interface board – HEIDENHAIN PWM 8 User Manual

Page 41

background image

-

41

-

Measure amplitudes:
Measuring range:

0.2 Vpp - 1.6 Vpp

Measuring frequency:

10 Hz - 50 kHz

Tolerance:
without adjustment via software:

±

5 %

with adjustment via software:

±

3 %


Terminating resistance:

121


Fault detection signal:

A and B < 0.3 Vpp


Encoder output:
Output

signal:

like

input

signal

with U

0


8.6 Specifications of TTL Interface Board


Maximum input voltage:

±

7V

Maximum input frequency:

approx. 2 MHz


Note:
The maximum input frequency only stands for the cutoff frequency of the square-wave input at PWM 8
(signal source: frequency generator).

Measure current/voltage:
Measuring range

current:

0 - 500 mA

Measuring range

voltage:

0 - 10 V

Tolerance:

±

3 %


Measure amplitudes:
Measuring range

high level:

2.5 - 7.5 V

Measuring range

low level:

0 - 2.5 V

Resolution:

50

mV

Measuring frequency:

10 Hz - 200 kHz

Tolerance:

±

50 mV


Terminating resistance:
From encoder signal to U-MSYS

215

From encoder signal to GND

90,9


Special feature of TTL interface board:
Owing to the input wiring, the PHA/TV display works properly even in the case of a cable break. The missing
signals are generated internally and output at the encoder output. Cable breaks can be found in the mode
"Measure Signal Amplitudes" or by checking the encoder signals at the BNC sockets.