6 specifications of ttl interface board – HEIDENHAIN PWM 8 User Manual
Page 41
-
41
-
Measure amplitudes:
Measuring range:
0.2 Vpp - 1.6 Vpp
Measuring frequency:
10 Hz - 50 kHz
Tolerance:
without adjustment via software:
±
5 %
with adjustment via software:
±
3 %
Terminating resistance:
121
Ω
Fault detection signal:
A and B < 0.3 Vpp
Encoder output:
Output
signal:
like
input
signal
with U
0
8.6 Specifications of TTL Interface Board
Maximum input voltage:
±
7V
Maximum input frequency:
approx. 2 MHz
Note:
The maximum input frequency only stands for the cutoff frequency of the square-wave input at PWM 8
(signal source: frequency generator).
Measure current/voltage:
Measuring range
current:
0 - 500 mA
Measuring range
voltage:
0 - 10 V
Tolerance:
±
3 %
Measure amplitudes:
Measuring range
high level:
2.5 - 7.5 V
Measuring range
low level:
0 - 2.5 V
Resolution:
50
mV
Measuring frequency:
10 Hz - 200 kHz
Tolerance:
±
50 mV
Terminating resistance:
From encoder signal to U-MSYS
215
Ω
From encoder signal to GND
90,9
Ω
Special feature of TTL interface board:
Owing to the input wiring, the PHA/TV display works properly even in the case of a cable break. The missing
signals are generated internally and output at the encoder output. Cable breaks can be found in the mode