beautypg.com

AMETEK MX CTSH User Manual

Page 89

background image

California Instruments

Revision H

89

User Manual

MX Series CTSH Compliance Test System


7.5

Test Setup

The following parameters and controls are available for executing the EN 61000-4-11 test:

Parameter

Description

Standard Revision

Select Edition 1.0 or Edition 2.0 as appropriate. The edition primarily affects
the voltage variation mode data entry method and the timing of the voltage
variations as indicated in

Figure 7-2

and

Figure 7-3

above.

Nominal Voltage

Opening the IEC 61000-4-11 test window will set this parameter to the
present steady state frequency setting of the AC source. If required, a new
value can be entered for use during the test. At the end of the test, the AC
source will remain at this frequency.

Nominal Frequency

Opening the IEC 61000-4-11 test window will set this parameter to the
present steady state voltage setting of the AC source. If required, a new
value can be entered for use during the test. At the end of the test, the AC
source will remain at this voltage.

Test Type

Available test types are Dips and Variations..

Phase Mode

The phase mode used is shown below the class selection. This mode
cannot be changed from this window. To change phase mode, close this
window and use the main front panel control window to select the desired
phase mode first.

Run

Starts the EN 61000-4-11 test using the parameters set by the user. Test
progress is shown at the bottom of the window. The sequence numbers
shown correspond to those in the illustration.

Abort

Aborts test in progress.

Figure 7-4: IEC61000-4-11 Test Setup screen

This manual is related to the following products: