beautypg.com

AMETEK MX CTSH User Manual

Page 19

background image

California Instruments

Revision H

19

User Manual

MX Series CTSH Compliance Test System


2.4.2

EUT Category Abbreviations used

The following connection types are supported using the abbreviations shown in the table:

Abbreviation

Meaning

CSPNBTPE

Connection Single Phase & Non Balanced Three Phase Equipment

CBTPE

Connection Balanced Three Phase Equipment

CBTPEUSC

Connection Balanced Three Phase Equipment Under Specified Conditions.

If different Sce values apply to each category EUT, they will be shown between brackets immediately
following these abbreviations in the Limit drop down control.

2.4.3

IEC 61000-3-12 Test Times

The test time used to run a Harmonics test is determined by the nature of the EUT. Under the test
standard, four test times (observation periods) are allowed depending on the nature of the EUT. All are
aimed at ensuring repeatability of test results when tests are performed under the same conditions and on
the same test system. Repeatability for this purpose is defined as results that are within 5 %. Available
observation periods are:

Table 2-5: Harmonics Test Times

Type of equipment behavior

Observation period

Quasi-stationary

T

obs

of sufficient duration to meet the requirements for

repeatability.

Short cycles (Tcycle < 2.5 min)

T

obs

> 10 cycles (reference method) or Tobs of sufficient

duration or synchronization to meet the requirements for
repeatability. 'Synchronization' means that the total
observation period is sufficiently close to including an exact
integral number of equipment cycles such that the
requirements for repeatability are met.

Random

T

obs

of sufficient duration to meet the requirements for

repeatability.

Long cyclic (Tcycle > 2.5 min)

Full equipment program cycle (reference method) or a
representative 2.5 min period considered by the manufacturer
as the operating period with the highest total harmonic
content.

You can increase the test time up to 24 hours if needed. The test period needs to be set long enough to
cover the entire operating cycle of the unit under test. This is particularly important for the transitory
harmonics tests. Avoid setting the test time longer than necessary as it only decreases your test
throughput.

This manual is related to the following products: