14. Customizing EN 61000-4-13 Test Parameters

14.1

About This Chapter

This chapters covers describes how to modify aspects of EN61000-4 immunity testing to accommodate
different product committee requirements.

The available 61000-4 tests implemented in the MXGUI program are based on the most recent officially
published test standard. Future changes in test levels can be accommodated without the need to upgrade
the MXGUI program. To allow for some level of customization, the MXGUI will accept different test
parameters that can be edited by the user.

14.2

IEC413.413 File

[IEC-1000-4-13]

EN 61000-4-13 section header
Parameters that apply only to Flat Top Curve

Fcurv_lev = 90

Flat Curve Level at 90%

Fcurv_dwe = 120

Flat Curve Dwell time in seconds

Fcurv_pau = 5

Flat Curve Pause in seconds
Parameters that apply only to Over Swing Curve

Oswcurv_dwe = 120

Over Swing Curve Dwell time in seconds

Oswcurv_pau = 5

Over Swing Curve Pause in seconds

Oswcurv3_lev = 6

Over Swing Curve of third harmonic level at 6%

Oswcurv3_phs = 180

Over Swing Curve of third harmonic phase test of 180 degrees

Oswcurv5_lev = 4

Over Swing Curve of fifth harmonic level at 4%

Oswcurv5_phs = 0

Over Swing Curve of fifth harmonic phase test of 0 degree
Parameters that apply to Frequency Sweep

Swefreq1_lev = 3

Sweep Frequency Level of first range at 3%

Swefreq2_lev = 10

Sweep Frequency Level of second range at 10%

Swefreq3_lev = 5

Sweep Frequency Level of third range at 5%

Swefreq4_lev = 5

Sweep Frequency Level of fourth range at 5%

Swefreq5_lev = 2

Sweep Frequency Level of fifth range at 2%

Swefreq_dwe = 300

Sweep Frequency Dwell time in seconds

Swefreq1_stp = 5

Sweep Frequency Step size of first range in Hz

Swefreq2_stp = 10

Sweep Frequency Step size of second range in Hz

Swefreq3_stp = 10

Sweep Frequency Step size of third range in Hz

Swefreq4_stp = 10

Sweep Frequency Step size of fourth range in Hz

Swefreq5_stp = 25

Sweep Frequency Step size of fifth range in Hz
Parameters that apply only to Individual Harmonics

Indharm2_lev = 3

Individual harmonics Level for nth (second) harmonic in %

Indharm2_phs = 0

Individual harmonics Phase for nth (second) harmonic in degree

Indharm_dwe = 5

Individual harmonics Dwell time in second

Indharm_pau = 5

Individual harmonics Pause in second
Parameters that apply only to Interharmonics

Meisterfreqx_lev=4

Harmonic level for meister curve where x = 1 through 5

Meisterfreqx_stp=4

Frequency step size for meister curve where x = 1 through 5

Meisterfreqx_dwe=4

Meister curve harmonics Dwell time in second

Meisterfreqx_pau=4

Meister curve harmonics Pause in second

Meisterfreqx_lev=4

Harmonic level for meister curve where x = 1 through 5
Parameters that apply only to Interharmonics

Interharm1_lev = 2.5

Interharmonics Level for first range at 2.5%

Interharm2_lev = 5

Interharmonics Level for second range at 5%

Interharm3_lev = 3.5

Interharmonics Level for third range at 3.5%

Interharm4_lev = 2

Interharmonics Level for fourth range at 2%

Interharm5_lev = 1.5

Interharmonics Level for fifth range at 1.5%

Interharm_dwe = 5

Interharmonics Dwell time in second

Interharm_pau = 1

Interharmonics Pause in second