14. Customizing EN 61000-4-13 Test Parameters
14.1
About This Chapter
This chapters covers describes how to modify aspects of EN61000-4 immunity testing to accommodate
different product committee requirements.
The available 61000-4 tests implemented in the MXGUI program are based on the most recent officially
published test standard. Future changes in test levels can be accommodated without the need to upgrade
the MXGUI program. To allow for some level of customization, the MXGUI will accept different test
parameters that can be edited by the user.
14.2
IEC413.413 File
[IEC-1000-4-13]
EN 61000-4-13 section header
Parameters that apply only to Flat Top Curve
Fcurv_lev = 90
Flat Curve Level at 90%
Fcurv_dwe = 120
Flat Curve Dwell time in seconds
Fcurv_pau = 5
Flat Curve Pause in seconds
Parameters that apply only to Over Swing Curve
Oswcurv_dwe = 120
Over Swing Curve Dwell time in seconds
Oswcurv_pau = 5
Over Swing Curve Pause in seconds
Oswcurv3_lev = 6
Over Swing Curve of third harmonic level at 6%
Oswcurv3_phs = 180
Over Swing Curve of third harmonic phase test of 180 degrees
Oswcurv5_lev = 4
Over Swing Curve of fifth harmonic level at 4%
Oswcurv5_phs = 0
Over Swing Curve of fifth harmonic phase test of 0 degree
Parameters that apply to Frequency Sweep
Swefreq1_lev = 3
Sweep Frequency Level of first range at 3%
Swefreq2_lev = 10
Sweep Frequency Level of second range at 10%
Swefreq3_lev = 5
Sweep Frequency Level of third range at 5%
Swefreq4_lev = 5
Sweep Frequency Level of fourth range at 5%
Swefreq5_lev = 2
Sweep Frequency Level of fifth range at 2%
Swefreq_dwe = 300
Sweep Frequency Dwell time in seconds
Swefreq1_stp = 5
Sweep Frequency Step size of first range in Hz
Swefreq2_stp = 10
Sweep Frequency Step size of second range in Hz
Swefreq3_stp = 10
Sweep Frequency Step size of third range in Hz
Swefreq4_stp = 10
Sweep Frequency Step size of fourth range in Hz
Swefreq5_stp = 25
Sweep Frequency Step size of fifth range in Hz
Parameters that apply only to Individual Harmonics
Indharm2_lev = 3
Individual harmonics Level for nth (second) harmonic in %
Indharm2_phs = 0
Individual harmonics Phase for nth (second) harmonic in degree
Indharm_dwe = 5
Individual harmonics Dwell time in second
Indharm_pau = 5
Individual harmonics Pause in second
Parameters that apply only to Interharmonics
Meisterfreqx_lev=4
Harmonic level for meister curve where x = 1 through 5
Meisterfreqx_stp=4
Frequency step size for meister curve where x = 1 through 5
Meisterfreqx_dwe=4
Meister curve harmonics Dwell time in second
Meisterfreqx_pau=4
Meister curve harmonics Pause in second
Meisterfreqx_lev=4
Harmonic level for meister curve where x = 1 through 5
Parameters that apply only to Interharmonics
Interharm1_lev = 2.5
Interharmonics Level for first range at 2.5%
Interharm2_lev = 5
Interharmonics Level for second range at 5%
Interharm3_lev = 3.5
Interharmonics Level for third range at 3.5%
Interharm4_lev = 2
Interharmonics Level for fourth range at 2%
Interharm5_lev = 1.5
Interharmonics Level for fifth range at 1.5%
Interharm_dwe = 5
Interharmonics Dwell time in second
Interharm_pau = 1
Interharmonics Pause in second