AMETEK MX CTSH User Manual
Page 142

User Manual
MX Series CTSH Compliance Test System
17.6
Installing New CTSMXH Calibration Data
If you have send in the MXCTSH system for calibration at a remote calibration lab or at California
Instruments, the PC normally used to run the CTS system will typically not have the new calibration
on it when you receive the PACS-3-75 and A/D card back. The new calibration data will be
received on a floppy disk or emailed to you as an alternative.
To install the new calibration data, proceed as follows:
1. Insert the calibration disk received into floppy drive A
2. Open your Windows Explorer and select drive A.
3. Copy the Ctsmxh_Calibration.cts file located in the floppy root directory
4. Paste this file to the C:\ root of the PC that operates the CTS system.
5. Run the CTSMXH program. Upon launch, the CTSMXH program will detect the calibration
data file, which should have a newer date and time stamp that the calibration data currently in
use. If so, the program will automatically read the new cal file and update its calibration
database.
6. You can verify that the new calibration data has been loaded by selecting the "View,
Calibration Info" menu.
You can save the calibration disk if needed although annual re-calibration of the system is
recommended.
California Instruments
Revision G
139