13.8

Source Requirements

The EN 61000-4-29 imposes certain requirements on the DC test generator used. Not all of these
requirements can be met with the MX Series AC/DC power source. The following restrictions apply:

Capability

Standard Requirement

MX Series

Output voltage range

0 - 360 Vdc

0 - 400 Vdc

Output voltage variations under load

< 5 %

< 1 %

Ripple content

< 1 % of output voltage

< 1 V rms

Rise and fall time into 100 Ohm resistive
load

between 1 and 50 usec

0.3 V/usec

Over/undershoot into 100 Ohm resistive
load

< 10 %

< 2 %

Output current, steady state

up to 25 A

25 A

Note: The use of a test generator with higher or lower voltage or current capability is allowed provided that
the other specifications are preserved. The test generator steady state power/current capability shall be at
least 20 % greater than the EUT power/current ratings.

This means that for many EUT's a 25 A capable generator is not needed. However, since the rise and fall
time requirements may not be met under all circumstances, this is a pre-compliance test only.

For full compliance test applications, the test generator must be tested for compliance with the test
generator requirements.