AMETEK MX CTSH User Manual
Page 50

California Instruments
Revision H
50
User Manual
MX Series CTSH Compliance Test System
Exacq Technologies Card
After installing the A/D card in an available PCI slot, you must configure the card for operation in the CTS.
Generally, the PCI card will be the only A/D card to be used in your PC, but in case there are more cards
present, contact the factory for details concerning multi-card configuration.
Card Reference numbers:
CI Designation
Exacq P/N
System
CI401PCI
CM2110
Single phase systems
CI403PCI
CM2210
Three phase systems
After installing the PCI card, turn the computer back on. When Microsoft Windows starts up, it should
discover the new Exacq A/D device.
Next, you need to test the A/D card in your computer, and configure the A/D card for use with the CTS
software. After the PC starts up, you will have a new Program Group, called Exacq. Start the application
called Exacq Control Center from this new Exacq program group on the Windows Start menu. See Figure
3-13 below.
Figure 3-16: Exacq Control Center Screen.