5.9
Printing Test Reports
The Report button allows you to print both on-screen graphs displays and current and voltage harmonics
test result data to an MS Word document. This provides a test report that displays the test setup, the Pass
or Fail result and a graph of the voltage and current waveform as well as the worst case values for each of
the current harmonics. The Report button is available at the end of a test run or any time a test is aborted
with the Stop button.
Information included on the harmonic test report will be a function of the selected test mode and EUT
connection stage. The following fields provide this information:
Heading:
Test category:
Selected Stage and EUT type.
Current Test Result Summary
Test Result
Overall Pass or Fail result
I-THD (pk%)
Current distortion and applicable limit (%)
PWHD (%)
Partial weighted harmonic current distortion and
applicable limit
Highest parameters recorded during the test.
V_RMS
EUT Voltage Line to Neutral
I_Peak
Maximum peak current recorded
I_Fund
Maximum fundamental current recorded (I1)
Power
Maximum power level recorded
I_RMS
Maximum RMS current recorded
Crest Factor
Power Factor
Harmonic Currents Table:
Harmonic number (#)
Only harmonics for which limits apply are shown.
Harms (Avg)
Average harmonics value (1.5 sec smoothing
filter applied)
100% Limit
Absolute current value of 100 % limit for each
harmonic
% of Limit
Percent of 100 % limit for each harmonic
Harms (Max)
Maximum harmonic value recorded during test.
150% Limit
Absolute current value of 150 % limit for each
harmonic
% of Limit
Percent of 150 % limit for each harmonic
maximum
Status
Individual harmonics pass or fail result (based on
100% and 150% limits)
Voltage Source Verification Data