MX Series CTSH Compliance Test System


5.9

Printing Test Reports

The Report button allows you to print both on-screen graphs displays and current and voltage harmonics
test result data to an MS Word document. This provides a test report that displays the test setup, the Pass
or Fail result and a graph of the voltage and current waveform as well as the worst case values for each of
the current harmonics. The Report button is available at the end of a test run or any time a test is aborted
with the Stop button.

Information included on the harmonic test report will be a function of the selected test mode and EUT
connection stage. The following fields provide this information:

Heading:

Test category:

Selected Stage and EUT type.

Current Test Result Summary

Test Result

Overall Pass or Fail result

I-THD (pk%)

Current distortion and applicable limit (%)

PWHD (%)

Partial weighted harmonic current distortion and
applicable limit

Highest parameters recorded during the test.

V_RMS

EUT Voltage Line to Neutral

I_Peak

Maximum peak current recorded

I_Fund

Maximum fundamental current recorded (I1)

Power

Maximum power level recorded

I_RMS

Maximum RMS current recorded

Crest Factor

Power Factor

Harmonic Currents Table:

Harmonic number (#)

Only harmonics for which limits apply are shown.

Harms (Avg)

Average harmonics value (1.5 sec smoothing
filter applied)

100% Limit

Absolute current value of 100 % limit for each
harmonic

% of Limit

Percent of 100 % limit for each harmonic

Harms (Max)

Maximum harmonic value recorded during test.

150% Limit

Absolute current value of 150 % limit for each
harmonic

% of Limit

Percent of 150 % limit for each harmonic
maximum

Status

Individual harmonics pass or fail result (based on
100% and 150% limits)

Voltage Source Verification Data