MX Series CTSH Compliance Test System


9.2

Test Setup

The following parameters and controls are available for executing the EN 61000-4-14 test:

Control / Field

Description

Nominal voltage

Opening this window will set this parameter to the present steady state
voltage setting of the AC source. If required, a new value can be
entered for use during the test. At the end of the test, the AC source will
remain at this voltage level. If needed, the program will switch the AC
source to the high voltage range and close the output relay when the
test is started.

Nominal frequency

Opening this window will set this parameter to the present steady state
frequency setting of the AC source. If required, a new value can be
entered for use during the test. At the end of the test, the AC source will
remain at this frequency.

Test Method

See under Test Options below

Source Regulation

See under Test Options below

EUT Class

The EUT class determines the test levels used. Class selection is done
by loading the correct class file from disk..

Phase Mode

The phase mode is determined by the AC source configuration. This
mode cannot be changed from this window. To change phase mode,
close this window and use the main front panel control window to select
the desired phase mode first.

Start

Starts the EN 61000-4-14 test using the parameters set by the user.
Test progress is shown at the bottom of the window. The sequence
numbers shown correspond to those in the illustration.

Stop

Aborts test in progress.