Cilxs_configuretriggeredfunction – AMETEK Ls Series Driver Manual User Manual
Page 92

IVI Driver Manual
Lx / Ls Series
88
ciLxs_ConfigureTriggeredFunction
ViStatus ciLxs_ConfigureTriggeredFunction (ViSession instrumentHandle,
ViChar _VI_FAR function[]);
Purpose
This function configures the shape of the output voltage waveform.
Parameter List
instrumentHandle
Variable Type ViSession
The ViSession handle that you obtain from the ciLxs_init or
ciLxs_InitWithOptions function. The handle identifies a particular
instrument session.
Default Value: None
function
Variable Type ViChar[]
Pass the shape of the output voltage waveform when a triggered step
or pulse transient occurs. The driver sets the
CILXS_ATTR_TRIGGERED_FUNCTION attribute to this value.
Valid Function Names:
"SINUSOID","SQUARE","CSINUSOID" and user defined functions
Default Value: "SINUSOID"
Return Value
Returns the status code of this operation. The status code either
indicates success or describes an error or warning condition. You
examine the status code from each call to an instrument driver
function to determine if an error occurred.
To obtain a text description of the status code, call the
ciLxs_error_message function. To obtain additional information about
the error condition, call the ciLxs_GetError function. To clear the
error information from the driver, call the ciLxs_ClearError
function.
The general meaning of the status code is as follows:
Value Meaning
-------------------------------
0 Success
Positive Values Warnings