Cilxs_querymaxvoltagelevel – AMETEK Ls Series Driver Manual User Manual

Page 218

background image

IVI Driver Manual

Lx / Ls Series

214

ciLxs_QueryMaxVoltageLevel


ViStatus ciLxs_QueryMaxVoltageLevel (ViSession instrumentHandle,
ViPReal64 maxVoltageLevel);

Purpose

This function returns the maximum programmable voltage level that the
power supply accepts for a particular current limit on a phase for the
output range to which the power supply is currently configured.


Parameter List

instrumentHandle

Variable Type ViSession

The ViSession handle that you obtain from the ciLxs_init or
ciLxs_InitWithOptions function. The handle identifies a particular
instrument session.

Default Value: None


maxVoltageLevel

Variable Type ViReal64 (passed by reference)

This parameter returns the maximum programmable voltage level of the
AC source.

Units: volts (V)

Note:

1) This value is valid only for sine function.

Return Value

Returns the status code of this operation. The status code either
indicates success or describes an error or warning condition. You
examine the status code from each call to an instrument driver
function to determine if an error occurred.

To obtain a text description of the status code, call the
ciLxs_error_message function. To obtain additional information about
the error condition, call the ciLxs_GetError function. To clear the
error information from the driver, call the ciLxs_ClearError
function.

The general meaning of the status code is as follows:

Value Meaning
-------------------------------
0 Success
Positive Values Warnings