AMETEK Ls Series Driver Manual User Manual
Page 50

IVI Driver Manual
Lx / Ls Series
46
the CILXS_ATTR_CURRENT_LIMIT_BEHAVIOR attribute.
Defined Values:
CILXS_VAL_CURRENT_REGULATE - Regulatory limit
CILXS_VAL_CURRENT_TRIP - Trip limit
Default Value: CILXS_VAL_CURRENT_REGULATE
limit
Variable Type ViReal64
Pass the rms current limit of the specified output phase. The driver
uses this value to set the CILXS_ATTR_CURRENT_LIMIT attribute.
Units: amps
Valid Range (1 phase mode): 0.0 to 20.0 (3000iL)
0.0 to 30.0 (4500iL)
0.0 to 32.0 (4801iL)
Valid Range (3 phase mode): 0.0 to 6.7 (3000iL)
0.0 to 10.0 (4500iL)
Default Value: 0.0
Return Value
Returns the status code of this operation. The status code either
indicates success or describes an error or warning condition. You
examine the status code from each call to an instrument driver
function to determine if an error occurred.
To obtain a text description of the status code, call the
ciLxs_error_message function. To obtain additional information about
the error condition, call the ciLxs_GetError function. To clear the
error information from the driver, call the ciLxs_ClearError
function.
The general meaning of the status code is as follows:
Value Meaning
-------------------------------
0 Success
Positive Values Warnings
Negative Values Errors
This driver defines the following status codes:
Status Description
-------------------------------------------------
ERRORS:
BFFA1001 The trigger source is not software trigger.